Best practices for standardized performance testing of infrared thermographs intended for fever screening
Pejman Ghassemi1, T Joshua Pfefer1, Jon P Casamento1
1Center for Devices and Radiological Health, U.S. Food and Drug Administration, Silver Spring, Maryland, United States of America.
Abstract:
Infrared (IR) modalities represent the only currently viable mass fever screening approaches for outbreaks of infectious disease pandemics such as Ebola virus disease and severe acute respiratory syndrome. Non-contact IR thermometers (NCITs) and IR thermographs (IRTs) have been used for fever screening in public areas such as airports. While NCITs remain a more popular choice than IRTs, there has been increasing evidences in the literature that IRTs can provide great accuracy in estimating body temperature if qualified systems are used and appropriate procedures are consistently applied. In this study, we addressed the issue of IRT qualification by implementing and evaluating a battery of test methods for objective, quantitative assessment of IRT performance based on a recent international standard (IEC 80601-2-59). We tested two commercial IRTs to evaluate their stability and drift, image uniformity, minimum resolvable temperature difference, and radiometric temperature laboratory accuracy. Based on these tests, we illustrated how experimental and data processing procedures could affect results, and suggested methods for clarifying and optimizing test methods. Overall, the insights into thermograph standardization and acquisition methods provided by this study may improve the utility of IR thermography and aid in comparing IRT performance, thus improving the potential for producing high quality disease pandemic countermeasures.
More Related Videos
04:19A New Best Practice for Validating Tail Vein Injections in Rat with Near-infrared-Labeled Agents
Published on: April 19, 2019
11:46Investigation of the Electrophysiological and Thermographic Safety Parameters of Surgical Energy Devices During Thyroid and Parathyroid Surgery in a Porcine Model
Published on: October 13, 2022
Related Concept Videos
Testing a Claim about Standard Deviation
The hypothesis testing for the claim of population standard deviation (or variance) requires the data and samples to be random and unbiased. The population distribution also must be normal. There is no specific requirement on the sample size as the estimation is based on the chi-square distribution.
As a first step, the hypothesis (null and alternative) concerning the claim about...
Types of Fever
Here are the different types of fever:
Patterns of Fever
Methods of reducing fever
Pharmacological Methods of Reducing Fever:
Standard Enthalpy of Formation
Standard Electrode Potentials