Optimization of exposure parameters in dental cone beam computed tomography using a 3-step approach
Soontra Panmekiate1, Peera Rungwittayathon1, Wijuck Suptaweeponboon1
1Department of Radiology, Faculty of Dentistry, Chulalongkorn University, Bangkok, Thailand.
Objective:
The aim of this study was to find the optimal balance among tube voltage (kV), tube current (mA), and exposure time (s) in cone beam computed tomography (CBCT).
Study Design:
Three human hemimandibles were scanned by using the 3D Accuitomo 170 scanner (J. Morita, Kyoto, Japan). First, 3 combinations of kilovolt (kV) and milliampere (mA) were used at a constant radiation dose. Seven observers evaluated the images; the kV of the highest-scoring scan was considered optimal. Second, the lowest acceptable mA for visualizing different anatomic structures was determined. Finally, the samples were scanned by using 3 combinations of tube current and exposure time; the observers determined the combination with the highest image quality.
Results:
At a constant radiation dose, the highest available voltage (i.e., 90 kV) resulted in the highest image quality in terms of general impression, sharpness, noise, and artefacts. Depending on the anatomic structure, mA reductions of 20% to 40% compared with the default setting were possible. Fast-scan protocols showed equal or slightly better image quality compared with the standard-scan mode.
Conclusions:
For the CBCT model used in this study, optimization implies the use of the highest kV along with the shortest exposure time and a task-specific mA. The proposed stepwise optimization approach could be applied to any CBCT unit, preferably during commissioning.
More Related Videos
10:23Author Spotlight: Three-Dimensional Cephalometric Landmark Annotation Demonstration on Human Cone Beam Computed Tomography Scans
Published on: September 8, 2023
05:37Cone Beam Intraoperative Computed Tomography-based Image Guidance for Minimally Invasive Transforaminal Interbody Fusion
Published on: August 6, 2019
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies III: Computed Tomography
Model Approaches for Pharmacokinetic Data: Distributed Parameter Models
The distributed parameter models are specifically designed to account for variations and differences in some drug classes. This model is particularly useful for assessing regional concentrations of anticancer or...
Beams
Based on geometry, beams can be straight, tapered, or curved. Straight beams are the most common type and have a constant cross-section throughout their length. Tapered beams, on the other hand, have a varying cross-section along...
Deflection of a Beam
Singularity functions, described in an earlier lesson, are powerful mathematical tools that represent discontinuities within a function commonly encountered in structural loading...
Wave Parameters