Related Experiment Video
Updated: Feb 5, 2026

Discovering Middle Ear Anatomy by Transcanal Endoscopic Ear Surgery: A Dissection Manual
Published on: January 11, 2018
Case Report of a New Coupler for Round Window Application of an Active Middle Ear Implant
Thomas Lenarz1,2, Denise Zimmermann1, Hannes Maier1,2
1Department of Otorhinolaryngology, Hannover Medical School.
Objectives:
To evaluate feasibility, surgical handling, audiological outcome, and coupling efficiency of a new coupler (custom-made device) for an active middle ear implant.
Patient:
Revision surgery after implantation of an active middle ear implant in a 66-year-old male patient with mixed hearing loss.
Intervention:
Prosthetic hearing rehabilitation with a new coupler for round window application.
Main Outcome And Results:
The patient obtained good speech perception in quiet (word recognition scores 80%; Freiburg monosyllables) and noise (-3.3 dB SNR; Oldenburg Sentence Test). The effective gain with the Hannover coupler improved at frequencies > 0.5 kHz compared with the values reported for other round window (RW)-coupling modalities.
Conclusion:
The coupler provides a feasible option for the RW application of the middle ear implant actuator. The spring concept of the coupler needs to be improved to further standardize RW-coupling and improve coupling efficiency at low frequencies (0.5 kHz).
Related Concept Videos
Anatomy of the Ear
Data Reporting and Recording
Types of Reports I: Hands-off Report
Following are the key components and categories of hand-off reports:
Purpose and Process:
Types of Reports II: Incident or Occurrence Report
Purposes:
In the healthcare industry, reports play a crucial role in documenting incidents within an agency. The primary objective of these reports is to ensure patient safety, uphold the...
Types of Reports III: Telephone and Verbal Reports
Here's an overview of each type:
Telephone Orders
Reporter Genes