Improving image quality with model-based iterative reconstruction at quarter of nominal dose in upper abdominal CT
Xirong Zhang1,2, Jing Chen2, Nan Yu2
1Department of Medical Techniques, Shaanxi University of Chinese medicine, Xianyang, China.
Objective:
To evaluate the ability of a model-based iterative reconstruction (MBIR) for improving image quality in upper abdominal CT with quarter of the normal dose, in comparison with adaptive statistical iterative reconstruction (ASiR) at normal dose.
Methods:
40 upper abdominal patients were randomly divided into two groups: normal-dose group (n = 20) with tube current modulation for noise index (NI) of 10 HU and 40% ASiR reconstruction; low-dose group (n = 20) with NI = 20 HU in the delay phase and MBIR and 40%ASiR. Images in the delay phase were compared. The CT values and standard deviation (SD) values of the liver, spleen, pancreas, kidney, erector spine and fat were measured. Contrast-noise-ratio (CNR = (CTtissue-CT fat)/SDfat) of each measured organ were calculated and compared with one-way ANOVA among the three reconstruction groups. The subjective image scores of the three groups were assessed blindly by two experienced physicians using a 5-point system and the score consistency was compared by the κ test.
Results:
Dose reduction of 75 % was achieved for the low-dose scan. The subjective scores (95 % confidence intervals) of the three groups (NI 10-40 % ASiR, NI 20-40% ASiR and NI 20-MBIR) were 4.00 ± 0.79 (3.62-4.37), 3.35 ± 0.58 (3.07-3.62) and 3.90 ± 0.64 (3.60-4.19), respectively with no difference between the NI 10-40% ASiR and NI20-MBIR groups and good consistency between reviewers (κ = 0.726). MBIR had statistically lower SD values and higher contrast-to-noise ratio values in the liver, spleen, pancreas, kidney and erector spine than NI 10-40% ASiR and NI 20-40% ASiR (all p < 0.05).
Conclusion:
At 75 % dose reduction, MBIR provides similar image quality compared to 40% ASiR at normal-dose.
Advances In Knowledge:
MBIR provides good image quality at 25 % of the normal dose.
More Related Videos
Related Concept Videos
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Imaging Studies for Cardiovascular System V: CT
Transformers with Off-Nominal Turns Ratios
Nominal Level of Measurement
The data that cannot be measured but can be grouped into categories fall under the nominal level of measurement. Data that is measured using a nominal...
Sign Test for Nominal Data
For example, consider a...
Imaging Studies for Cardiovascular System VI: Calcium -Scoring CT