In-vitro validation of 4D flow MRI measurements with an experimental pulsatile flow model
A David1, D Le Touze2, K Warin-Fresse1
1Department of cardiovascular imaging, centre hospitalier universitaire de Nantes, 44800 Saint-Herblain, France.
Purpose:
The purpose of this study was to assess the precision of four-dimensional (4D) phase-contrast magnetic resonance imaging (PCMRI) to measure mean flow and peak velocity (Vmax) in a pulsatile flow phantom and to test its sensitivity to spatial resolution and Venc.
Material And Methods:
The pulsatile flow phantom consisted of a straight tube connected to the systemic circulation of an experimental mock circulatory system. Four-dimensional-PCMR images were acquired using different spatial resolutions (minimum pixel size: 1.5×1.5×1.5mm3) and velocity encoding sensitivities (up to three times Vmax). Mean flow and Vmax calculated from 4D-PCMRI were compared respectively to the reference phantom flow parameters and to Vmax obtained from two-dimensional (2D)-PCMRI.
Results:
4D-PCI measured mean flow with a precision of -0.04% to+5.46%, but slightly underestimated Vmax when compared to 2D-PCMRI (differences ranging from -1.71% to -3.85%). 4D PCMRI mean flow measurement was influenced by spatial resolution (P<0.001) with better results obtained with smaller voxel size. There was no effect of Venc on mean flow measurement. Regarding Vmax, neither spatial resolution nor Venc did influence the precision of the measurement.
Conclusion:
Using an experimental pulsatile flow model 4D-PCMRI is accurate to measure mean flow and Vmax with better results obtained with higher spatial resolution. We also show that Venc up to 3 times higher than Vmax may be used with no effect on these measurements.
More Related Videos
10:02Characterization of the Isolated, Ventilated, and Instrumented Mouse Lung Perfused with Pulsatile Flow
Published on: April 29, 2011
07:30In Vitro Model of Physiological and Pathological Blood Flow with Application to Investigations of Vascular Cell Remodeling
Published on: November 3, 2015
Related Concept Videos
Gene Flow
Flow Cytometry
In...
Flow Sheet
Here's a closer look at the examples of flowsheets commonly used by nurses:
Graphic Sheet Documentation:
Flow Table Test
Concrete is placed within a truncated cone mold that is 8 inches high with an 8-inch base diameter and a 5-inch top diameter. The...
Irrotational Flow
Turbulent Flow