[Determination of p-Anisidine in workplace air by HPLC]
1Henan Provincial Institute of Occupational Health, Zhengzhou 450052, China.
Abstract:
Objective: To establish an HPLC method for determining p-Anisidine in workplace air. Methods: An C(18) capillary column was used with the detection wavelength of 240 nm, the methanol solution (v/v=62:38) as mobile phase, the flow rate was 0.9 ml/min, and the injection volume was 20 μl. Results: The linear range (μg/ml) for p-Anisidine was 0~100; correlation coefficient was 0.999 7; the detection limit was 0.15 μg/ml; the average precision of RSD was 4.4%. The sampling efficiency was more than 95% using silicone tube as sampling filter. Sample can be stored at room temperature for more than 7 d. Conclusion: the method appered to be accurate, simple, sensitive and reliable, also have a high sampling efficiency, and can be usd to determine p-anisidine in workplace air simultaneously.