A Novel Rescue Technique for Difficult Intubation and Difficult Ventilation
Published on: January 17, 2011
The Channelled Airtraq® as a Rescue Device Following Failed Expected Difficult Intubation with an Angulated Video
1Department of Anaesthesiology and Reanimation, Kocaeli University School of Medicine, Kocaeli, Turkey.
Abstract:
We herein report two expected difficult intubation cases that failed with an angulated-type video laryngoscope (C-MAC D-blade) and were rescued with a channelled-type Airtraq® device. The common signs and characteristics which indicated difficult ventilation and intubation in these patients were Mallampati 4 (with phonation), mandibular protrusion of B, obstructive sleep apnoea disorder, male gender, and thick neck (>46 cm). We had aids ready in the operating theatre for the anticipated difficult intubation. We first attempted to intubate the trachea in two patients with direct laryngoscopy; as expected, the Cormack-Lehane (CL) grades of the two patients were 4, even cricoid pressure was applied. Second, we attempted to intubate with the angulated-type C-MAC D-blade; the CL grades improved to 2. However, despite tube adjustment manoeuvres and use of a rigid stylet, we were unable to insert the tube into the trachea. Then, we attempted to intubate with a channelled-type Airtraq® device. Consequently, without need for a stylet or use of any manoeuvres, we were able to intubate the tracheas at the first attempt.
More Related Videos
03:58Author Spotlight: Advancing Awake Nasotracheal Intubation with Flexible Video Rhino-Laryngoscopes
Published on: August 2, 2024
05:39Author Spotlight: A Non-Intubated Video-Assisted Thoracoscopic Surgery with Multimodal Analgesia and Sevoflurane Inhalation Anesthesia
Published on: May 26, 2023
Related Concept Videos
Expected Value
Hypothesis: Accept or Fail to Reject?
There are two ways to indicate that the null hypothesis is not rejected. 'Accept' the null...
Determination of Expected Frequency
Ion Channels
Ion channels are specialized integral membrane proteins on the plasma membrane that allow...
Endotracheal Intubation I: Procedure
The ET tube comprises various components, including a standard adaptor to attach a bag-valve-mask (BVM) or ventilator, a cuff, a pilot balloon, and radiopaque markings along its length to measure the insertion distance. The tube sizes...
Expected Frequencies in Goodness-of-Fit Tests