Real-Time Measurement of Xenon Concentration in a Binary Gas Mixture Using a Modified Ultrasonic Time-of-Flight
David J Williams1,2, Gregory D Hallewell3, Ela Chakkarapani4
1From the Department of Anaesthetics, Morriston Hospital, Swansea, United Kingdom.
Background:
Xenon (Xe) is an anesthetic gas licensed for use in some countries. Fractional concentrations (%) of gases in a Xe:oxygen (O2) mixture are typically measured using a thermal conductivity meter and fuel cell, respectively. Speed of sound in such a binary gas mixture is related to fractional concentration, temperature, pressure, and molar masses of the component gases. We therefore performed a study to assess the feasibility of developing a novel single sterilizable device that uses ultrasound time-of-flight to measure both real-time flowmetry and fractional gas concentration of Xe in O2.
Methods:
For the purposes of the feasibility study, we adapted an ultrasonic time-of-flight flowmeter from a conventional anesthetic machine to additionally measure real-time fractional concentration of Xe in O2. A total of 5095 readings of Xe % were taken in the range 5%-95%, and compared with simultaneous measurements from the gold standard of a commercially available thermal conductivity Xe analyzer.
Results:
Ultrasonic measurements of Xe (%) showed agreement with thermal conductivity meter measurements, but there was marked discontinuity in the middle of the measurement range. Bland-Altman analysis (95% confidence interval in parentheses) yielded: mean difference (bias) 3.1% (2.9%-3.2%); lower 95% limit of agreement -4.6% (-4.8% to -4.4%); and upper 95% limit of agreement 10.8% (10.5%-11.0%).
Conclusions:
The adapted ultrasonic flowmeter estimated Xe (%), but the level of accuracy is insufficient for clinical use. With further work, it may be possible to develop a device to perform both flowmetry and binary gas concentration measurement to a clinically acceptable degree of accuracy.
More Related Videos
10:45Simultaneous Electroencephalography, Real-time Measurement of Lactate Concentration and Optogenetic Manipulation of Neuronal Activity in the Rodent Cerebral Cortex
Published on: December 19, 2012
09:50Technical Aspect of the Automated Synthesis and Real-Time Kinetic Evaluation of [11C]SNAP-7941
Published on: April 28, 2019
Related Concept Videos
Kinetic Molecular Theory and Gas Laws Explain Properties of Gas Molecules
Gas Exchange and Transport
Gas Laws: Boyle's, Gay-Lussac, Charles', Avogadro's, and Ideal Gas Law
Drug Concentration Versus Time Correlation
Two pivotal parameters are the minimum effective concentration (MEC) and the minimum toxic concentration (MTC). The MEC is the...
The Integrated Rate Law: The Dependence of Concentration on Time
Behavior of Gas Molecules: Molecular Diffusion, Mean Free Path, and Effusion