Related Experiment Video
Updated: Feb 4, 2026

A Murine Tail Lymphedema Model
Published on: February 10, 2021
Electrode Equivalence for Use in Bioimpedance Spectroscopy Assessment of Lymphedema
Brenda J Svensson1,2, Elizabeth S Dylke1, Leigh C Ward1,3
11 Faculty of Health Sciences, The University of Sydney, Sydney, Australia.
Background:
Bioimpedance spectroscopy (BIS) is commonly used in the assessment and monitoring of lymphedema. This study investigated electrodes as a source of variability that could impact the accuracy of BIS in the clinic and determined if Ag/AgCl electrocardiograph (ECG) electrodes could be used as an alternative to instrument-specific electrodes.
Methods And Results:
Two types of Ag/AgCl electrodes were studied: instrument-specific bioimpedance electrodes (bioimpedance) and single tab ECG electrodes (cardiac). Six areas of investigation were addressed: intrinsic electrode resistance; electrode age; drive electrode position; electrode width/surface area; concordance between cardiac and bioimpedance electrodes; and mixing electrode types and batches. Participants included women (n = 26) and men (n = 8), both with (n = 4) and without lymphedema (n = 30). Resistance (R0) of the limbs was measured and used to calculate interlimb BIS ratios. Intrinsic electrode resistance varied between batches (p ≤ 0.001), with cardiac electrodes recording higher resistance. Electrode age had no impact on limb resistance (p = 0.85). Drive electrode position biased limb resistance (0.1%-2.3%) and electrode size/surface area had a small (≤1%), but significant effect on limb resistance (p ≤ 0.001). However, calculation of interlimb BIS ratios negated the impact of these as well as any effect of mixing electrode batches and types (p = 0.15-0.96). Electrode type had no impact on arm and leg resistance, or interlimb BIS ratios (p = 0.173-0.289).
Conclusion:
Calculation of interlimb BIS ratios improves accuracy of clinical BIS. Ag/AgCl cardiac electrodes can be used as an alternative to device-specific electrodes to measure limb resistance.
Related Concept Videos
Equivalent Capacitance
The following strategies are adopted to calculate...
Equivalent Capacitance
Pharmaceutical Equivalents
Thévenin Equivalent Circuits
Principle of Equivalence
Equivalent Couples
Two couples are considered to be equivalent if they produce the same rotational effect on a rigid body. In other words, the two couples have the same magnitude and act in the same direction, causing the same angular displacement or acceleration in the body.
For instance, consider two couples lying in the plane of the page, with one having a pair of equal...