Biomechanical Assessment of Torsional Stiffness in a Supracondylar Humerus Fracture Model
Melissa Wallace1, David B Johnson2, William Pierce1
1Department of Orthopaedic Surgery, Texas Scottish Rite Hospital, Dallas, TX.
Background:
We assessed the effect on the torsional stability by different pin diameters and varied pin configurations in a biomechanical supracondylar humerus fracture model.
Methods:
After scanning a model of a pediatric humerus, the image was imported into software. Variable pin trajectories were planned. Acrylonitrile butadiene styrene plastic models were 3-dimensionally printed with predetermined pin trajectories. Models were osteotomized and potted with a polyurethane resin. Five-pin configurations were designed to test coronal and sagittal patterns of pin placement. Each included 3 lateral pins and a medial pin. Pin diameters of 1.6, 2.0, and 2.4 mm were tested in all configurations. Three models for each pin diameter/configuration were tested to ensure uniformity. Stability of the construct was tested to determine the torque needed to deflect the osteotomy 10 degrees in internal/external rotation. Each model was tested 3 times.
Results:
In all models/configurations, the 2.4 mm pin diameter was statistically stiffer than 1.6 mm diameter pins; this lost statistical significance in certain patterns when comparing 2.0- and 2.4-mm pins. When comparing a divergent to a parallel configuration in the coronal plane, there was no significant difference in stability when pin diameter or number were controlled. The convergent pin configuration was, in general, the least stable pattern. Use of a medial pin conferred statistically significant stiffness throughout most models as demonstrated with pin deletion. Use of 2 pins was significantly less stiff than most 3-pin models.
Conclusions:
Larger pin diameters confer greater stiffness among all patterns. The use of 3 lateral and 1 medial pin was not statistically different than 2 lateral and 1 medial pin in our models. Both patterns were stiffer than 3 lateral pins only or other fewer pin constructs. The alignment of pins in the sagittal plane did not affect overall construct stiffness.
Related Concept Videos
Torsion of Noncircular Members
Torsional Pendulum
As long as the rigid body's angular displacement is small, its oscillation can be modeled as a linear angular oscillation. The amplitude of the oscillation is an angle. The role of mass is played...
Bending and Torsional Moments
The reaction developed in a structural element when subjected to an external force causes the element to bend. When a structural element bends upwards, it creates compressive normal forces on the top and tensile normal forces on the bottom, resulting in a couple that determines the bending...
Bones of the Upper Limb: Humerus
Fractures: Bone Repair
Minor fractures with no bone displacement are treated by immobilizing the fractured bone using a cast or splint. However, in the case of fractures with displaced bones, the broken bones are repositioned before immobilization to ensure successful healing without deformation and loss of function. The realignment of fractured bone ends is performed through a process called reduction. If the...
Assessment of the Mouth
Mouth Inspection
The inspection begins with visually examining the mouth for symmetry, color, and size.