On the accuracy and precision of PLANET for multiparametric MRI using phase-cycled bSSFP imaging
Yulia Shcherbakova1, Cornelis A T van den Berg2, Chrit T W Moonen1
1Center for Image Sciences, Imaging Division, University Medical Center Utrecht, Utrecht, the Netherlands.
Purpose:
In this work we demonstrate how sequence parameter settings influence the accuracy and precision in T1 , T2 , and off-resonance maps obtained with the PLANET method for a single-component signal model. In addition, the performance of the method for the particular case of a two-component relaxation model for white matter tissue was assessed.
Methods:
Numerical simulations were performed to investigate the influence of sequence parameter settings on the accuracy and precision in the estimated parameters for a single-component model, as well as for a two-component white matter model. Phantom and in vivo experiments were performed for validation. In addition, the effects of Gibbs ringing were investigated.
Results:
By making a proper choice for sequence parameter settings, accurate and precise parameter estimation can be achieved for a single-component signal model over a wide range of relaxation times at realistic SNR levels. Due to the presence of a second myelin-related signal component in white matter, an underestimation of approximately 30% in T1 and T2 was observed, predicted by simulations and confirmed by measurements. Gibbs ringing artifacts correction improved the precision and accuracy of the parameter estimates.
Conclusion:
For a single-component signal model there is a broad "sweet spot" of sequence parameter combinations for which a high accuracy and precision in the parameter estimates is achieved over a wide range of relaxation times. For a multicomponent signal model, the single-component PLANET reconstruction results in systematic errors in the parameter estimates as expected.
Related Concept Videos
Uncertainty in Measurement: Accuracy and Precision
Accuracy and Precision
The Water Cycle
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Acceleration due to Gravity on Other Planets
Astronomical observations are thus used to measure the acceleration due to gravity on other planets. This can be determined by observing the effect of a planet's gravity on objects close to it. The crucial factor that helps in this...
Improving Translational Accuracy