First Look at Control-IQ: A New-Generation Automated Insulin Delivery System
Sue Brown1,2, Dan Raghinaru3, Emma Emory1
1Center for Diabetes Technology, University of Virginia, Charlottesville, VA.
Objective:
To pilot test a new closed-loop control technology to validate it for a further large clinical trial.
Research Design And Methods:
The t:slim X2 insulin pump with Control-IQ Technology (Tandem Diabetes Care) includes a Dexcom G6 sensor and a closed-loop algorithm implemented in the pump that 1) automates insulin correction boluses, 2) has a dedicated hypoglycemia safety system, and 3) gradually intensifies control overnight, aiming for blood glucose levels of approximately 100-120 mg/dL every morning.
Results:
Five patients with type 1 diabetes (mean age 52.8 years, 2/3 male/female, mean A1C 6.5%) used Control-IQ in an outpatient setting (hotel) for approximately 37 h. During the closed loop, mean glucose was 129 mg/dL (135/121 mg/dL during the day/night), time within target range 70-180 mg/dL was 87% (82%/94% during the day/night), and time <60 mg/dL was 1.1% (2.0%/0.0% during the day/night).
Conclusions:
Following this pilot trial, Control-IQ was deployed in several studies, including the large-scale National Institutes of Health International Diabetes Closed-Loop (iDCL) Trial.
Related Concept Videos
Insulin Formulations: Types and Delivery
Short-acting insulins are divided into...
Generator Voltage Control
Insulin Secretory Vesicles
Insulin: The Receptor and Signaling Pathways
Insulin: Biosynthesis, Chemistry, and Preparation
Damage or functional impairment of β-cells inhibits insulin production, leading to diabetes. Diabetes treatment...
Insulin: Dosing Regimen and Adverse Effects
The basal dose constitutes about 40%-50% of the total daily dose, with the rest as premeal insulin. The mealtime insulin dose should mirror...