Multi-Wavelength Photoplethysmography Enabling Continuous Blood Pressure Measurement With Compact Wearable
Objective:
To fight the "silent killer" hypertension, continuous blood pressure (BP) monitoring has been one of the most desired functions in wearable electronics. However, current BP measuring principles and protocols either involve a vessel occlusion process with a cuff or require multiple sensing nodes on the body, which makes it difficult to implement them in compact wearable electronics like smartwatches and wristbands with long-term wearability.
Methods:
In this work, we proposed a highly compact multi-wavelength photoplethysmography (MWPPG) module and a depth-resolved MWPPG approach for continuous monitoring of BP and systemic vascular resistance (SVR). By associating the wavelength-dependent light penetration depth in the skin with skin vasculatures, our method exploited the pulse transit time (PTT) on skin arterioles for tracking SVR (n = 20). Then, we developed an arteriolar PTT-based method for beat-to-beat BP measurement. The BP estimation accuracy of the proposed arteriolar PTT method was validated against Finometer (n = 20) and the arterial line (n = 4).
Results:
The correlation between arteriolar PTT and SVR was theoretically deduced and experimentally validated on 20 human subjects performing various maneuvers. The proposed arteriolar PTT-based method outperformed the traditional arterial PTT-based method with better BP estimation accuracy and simpler measurement setup, i.e., with a single sensing node.
Conclusion:
The proposed depth-resolved MWPPG method can provide accurate measurements of SVR and BP, which are traditionally difficult to measure in a noninvasive or continuous fashion.
Significance:
This MWPPG work provides the wearable healthcare electronics of compact size with a low-cost and physiology-based solution for continuous measurement of BP and SVR.
Related Concept Videos
Measurement of Blood Pressure
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
The de Broglie Wavelength
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Blood Pressure
Blood Pressure
The average BP in an adult is typically around 120/80 mmHg (millimeters of mercury). In this measurement, the numerator (120) indicates the systolic pressure, which is the pressure in the arteries during the contraction of the heart's ventricles as blood is expelled. The denominator (80) represents the...