Aims:
To report the findings of an audit for radiotherapy errors from a low-middle-income country (LMICs) centre. This would serve as baseline data for radiotherapy error rates, their severity and causes, in such centres where modern error reporting and learning processes still do not exist.
Materials And Methods:
A planned cross-sectional weekly audit of electronic radiotherapy charts at the radiotherapy planning and delivery step for all patients treated with curative intent was conducted. Detailed analysis was carried out to determine the step of origin of error, time and contributing factors. They were graded as per indigenous institutional (TMC) radiotherapy error grading (TREG) system and the contributing factors identified were prioritised using the product of frequency, severity and ease of detection.
Results:
In total, 1005 consecutive radically treated patients' charts were audited, 67 radiotherapy errors affecting 60 patients, including 42 incidents and 25 near-misses were identified. Transcriptional errors (29%) were the most common type. Most errors occurred at the time of treatment planning (59.7%), with "plan information transfer to the radiation oncology information system" being the most frequently affected sub-step of the radiotherapy process (47.8%). More errors were noted at cobalt units (52/67; 77.6%) than at linear accelerators. Trend analysis showed an increased number of radiotherapy incidents on Fridays and near-misses on Mondays. Trend for increased radiotherapy errors noted in the evening over other shifts. On severity grading, most of the errors (54/60; 90%) were clinically insignificant (grade I/II). Inadequacies and non-adherence towards standard operating procedures, poor documentation and lack of continuing education were the three most prominent causes.
Conclusion:
Preliminary data suggest a vulnerability of LMIC set-up to radiotherapy errors and emphasises the need for the development of longitudinal prospective processes, such as voluntary reporting and a continued education system, to ensure robust and comprehensive safe practises on par with centres in developed countries.
Related Concept Videos
Systematic Error: Methodological and Sampling Errors
Sampling errors originate from improper sampling methods or the wrong sample population. These errors can be minimized by refining the sampling strategy. Defective instruments or faulty calibrations are the sources of instrumental...
Fundamental Attribution Error
Random Error
Margin of Error
Sampling Plans
Random sampling is a method where each member of the population has an equal chance of being selected for the sample. It involves selecting individuals randomly, often using random number generators or lottery-type methods. For example, when analyzing the properties of a...
Standard Error of the Mean
The standard error of the mean is an example of a standard error. It is a unique standard deviation known as the standard deviation of the sampling distribution of the mean. The standard error of the mean is a statistic that calculates how correctly a sample distribution represents a...