Related Experiment Video
Updated: Aug 3, 2026

14:31
Bioelectric Analyses of an Osseointegrated Intelligent Implant Design System for Amputees
Published on: July 15, 2009
Implanter's Integrated Information (I3) System: An Aid for Spinal Cord Stimulation Procedures
Kaare Meier1,2,3, Anne Lene Høst Knudsen1, Shellie Ann Boudreau4
1Department of Neurosurgery, Aarhus University Hospital, Aarhus, Denmark.
Summary
A new system aids spinal cord stimulation (SCS) lead implantation by integrating patient paresthesia descriptions and programming data. This visual dashboard improves the accuracy of paresthesia mapping for better pain relief.
Area of Science:
- Neurosurgery
- Biomedical Engineering
- Pain Management
Background:
- Spinal cord stimulation (SCS) uses electrical leads to evoke paresthesias for pain relief.
- Accurate paresthesia coverage of painful areas is crucial during SCS lead implantation.
- Patient descriptions of paresthesias are often imprecise and difficult to interpret.
Purpose of the Study:
- To optimize SCS implantation by improving patient paresthesia descriptions.
- To visually integrate lead position, programming parameters, and evoked paresthesias for physicians.
- To develop a user-friendly system for complex SCS data visualization.
Main Methods:
- Utilized a tablet app for patients to create detailed pain drawings and paresthesia mapping.
- Employed screen-sharing technology (Chromecast, Apple TV) to project data to multiple monitors.
- Positioned monitors adjacent to the fluoroscopy screen for real-time physician access.
Main Results:
- A three-monitor system provided a dynamic, integrated view of SCS implantation data.
- The system visualized lead position, programming parameters, and patient-reported paresthesias.
- This created an intuitive 'dashboard' for the implanting physician.
Conclusions:
- Introduced the Implanter's Integrated Information (I3) system for SCS procedures.
- The I3 system offers a simple, cost-effective solution for data integration and visualization.
- This approach enhances the precision and success rate of SCS lead implantation.