Related Experiment Video
Updated: Feb 3, 2026

Three-Dimensional Printing Guide Template Assisted Percutaneous Vertebroplasty PVP
Published on: October 17, 2019
Percutaneous MR-guided interventions using an optical Moiré Phase tracking system: Initial results.
Urte Kägebein1,2,3, Frank Godenschweger2, Brian S R Armstrong4
1Department of Diagnostic and Interventional Radiology, Hannover Medical School, Hannover, Germany.
This study developed real-time guidance for MR-guided interventions using optical Moiré Phase Tracking (MPT). The system provides accurate needle tip tracking and guidance, benefiting both expert and novice users in percutaneous procedures.
Area of Science:
- Medical Imaging
- Interventional Radiology
- Biomedical Engineering
Background:
- Percutaneous interventions require precise instrument navigation within Magnetic Resonance (MR) imaging.
- Real-time feedback is crucial for improving accuracy and safety during MR-guided procedures.
Purpose of the Study:
- To develop and evaluate a real-time optical Moiré Phase Tracking (MPT) guidance system for MR-guided percutaneous interventions.
- To assess the system's accuracy and usability for both experienced and novice interventional radiologists.
Main Methods:
- Modified a gradient echo sequence for real-time position updates using MPT markers attached to a needle.
- Developed an image guidance system displaying needle trajectory and tip-to-target distance as an overlay.
- Evaluated the system in a phantom study with 80 punctures on a 3T MRI system.
Main Results:
- The system achieved low mean errors: user error (0.91mm), system error (0.53mm), and total error (0.99mm).
- No significant differences in performance (time, error) were observed between expert and novice users.
- Continuous, interactive control of imaging planes and real-time feedback were enabled.
Conclusions:
- The developed MPT-based guidance system offers accurate and interactive real-time feedback for MR-guided percutaneous interventions.
- The system is effective for both experienced and novice users, potentially improving procedural safety and efficiency.
- This technology facilitates precise needle placement within the MR environment.
Related Concept Videos
Initiation of Translation
First, the initiator tRNA must be selected from the pool of elongator tRNAs by eukaryotic initiation factor 2 (eIF2). The initiator tRNA (Met-tRNAi) has conserved sequence elements including modified bases at...
Initiation of Translation
Nursing Interventions I: Taxonomy of Nursing Interventions
A nursing intervention is a treatment or action based on scientific concepts and knowledge from the nursing, behavioral, and physical sciences. Identifying and prioritizing nursing interventions based on the desired outcome...
Nursing Interventions II: Selecting and Classifying the Nursing Interventions
Transcription Initiation
The promoters and enhancers and their accessory proteins allow tight regulation of...
Phase Diagrams