A new method to statistically describe microcatheter tip position in patient-specific aneurysm models
Marie Teresa Nawka1, Jan-Hendrik Buhk1, Susanne Gellissen1
1Department of Diagnostic and Interventional Neuroradiology, University Medical Centre Hamburg-Eppendorf, Hamburg, Germany.
Background And Purpose:
Evidence on how to select microcatheters to facilitate aneurysm catheterization during coil embolization is sparse. We developed a new method to define microcatheter tip location inside a patient-specific aneurysm model as a 3-dimensional probability map. We hypothesized that precision and accuracy of microcatheter tip positioning depend on catheter tip shape and aneurysmal geometry.
Materials And Methods:
Under fluoroscopic guidance two to three operators introduced differently shaped microcatheters (straight, 45°, 90°) into eight aneurysm models targeting the anatomic center of the aneurysm. Each microcatheter position was recorded with flat-panel CT, and 3-dimensional probability maps of the microcatheter tip positions were generated. Maps were assessed with histogram analyses and compared between tip shapes, aneurysm locations and operators.
Results:
Among a total of 530 microcatheter insertions, the precision (mean distance between catheter positions) and accuracy (mean distance to target position) were significantly higher for the 45° tip (1.10±0.64 mm, 3.81±1.41 mm, respectively) than for the 90° tip (1.27±0.57 mm, p=0.010; 4.21±1.60 mm p=0.014, respectively). Accuracy was significantly higher in posterior communicating artery aneurysms (3.38±1.20 mm) than in aneurysms of the internal carotid artery (4.56±1.54 mm, p<0.001).
Conclusion:
Our method can be used tostatistically describe statistically microcatheter behavior in patient-specific anatomy, which may improve the available evidence guiding microcatheter shape selection. Experience increases the ability to reach the intended position with a microcatheter (accuracy) that is also dependent on the aneurysm location, whereas catheter tip choice determines the variability of catheter tip placements versus each other (precision). Clinical validation is required.
Related Concept Videos
Statistical Significance
Statistical Methods for Analyzing Epidemiological Data
Velocity and Position by Integral Method
Consider an example to calculate the velocity and position from the acceleration function. A motorboat is traveling at a constant velocity of 5.0 m/s when it starts to decelerate to arrive at the dock. Its acceleration is...
Velocity and Position by Graphical Method
Statistical Methods to Analyze Parametric Data: ANOVA
One-way ANOVA is applied when a single independent variable or factor is scrutinized. It compares...
Probability in Statistics
An example of a simple event is a coin toss. The result of a coin toss is either a head or a tail. Here, head and tail are two simple events. These two simple events make up the sample space. Further, the probability of an event occurring falls within the range of 0 to 1. The probability of an...