Repeatability, reproducibility and agreement of foveal avascular zone measurements using three different optical
Natasa Mihailovic1, Cristin Brand2, Larissa Lahme1
1Department of Ophthalmology, University of Muenster Medical Center, Muenster, Germany.
Purpose:
To evaluate the repeatability, the reproducibility and the agreement of foveal avascular zone (FAZ) measurements using three different optical coherence tomography angiography (OCT-A) devices.
Procedures:
This prospective study included 24 eyes of 24 healthy volunteers. OCT-A imaging was performed using RTVue XR Avanti, Canon OCT-HS100 and Spectralis OCT-A. Repeated measurements were performed under the same conditions on two separate days, and the area of the FAZ was determined and analyzed using the above devices.
Results:
All three devices showed a high ICC and there was no significant difference between the ICCs (pairwise comparison) of the three devices (Optovue-Canon (p = 0.66); Canon-Heidelberg (p = 0.21); Heidelberg-Optovue (p = 0.37). Agreement analysis of the three devices revealed a significant elevation of FAZ area values with the Heidelberg device and a slight underestimation of the FAZ area with the Canon device. Nevertheless, overall we found a high level of agreement between all of the three devices (ICC ≥ 0.958 (0.905-0.982)).
Conclusions:
Good reproducibility and repeatability were observed for all three devices. However, the agreement analysis revealed slight, but significant differences, which might limit alternating use of these devices for clinical research and follow-up examinations.
More Related Videos
Related Concept Videos
Hybrid Zones
Zones of Protection
Protective zones are defined by closed dashed lines, containing one or more components. A key characteristic of these zones is the strategic placement of...
Transition Zone
Bioequivalence Experimental Study Designs: Repeated Measures, Cross-Over, Carry-Over, and Latin Square Designs
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Non-ohmic Devices
Consider a simple circuit consisting of a battery, a diode, and a resistor. A...