Evaluation of optimization workflow using design of experiment (DoE) for various field configurations in
Kentaro Miki1, Akito Saito1, Takeo Nakashima2
1Department of Radiation Oncology, Hiroshima University Hospital, Japan.
Purpose:
In volumetric-modulated arc therapy (VMAT), field configurations such as couch or arc angles are defined manually or using a template. A field configuration is reselected through trial-and-error in the case of undesirable resultant planning. To efficiently plan for desirable quality, configurations should be assessed before dose calculation. Design of experiments (DoE) is an optimization technique that efficiently reveals the influence of inputs on outputs. We developed an original tool using DoE to determine the field configuration selection and evaluated the efficacy of this workflow for clinical practice.
Methods:
Computed-tomography scans of 17 patients and target structures were acquired retrospectively from a brain tumor treated using a dual-arc VMAT plan. The configurations of the couch, arc, collimator angles, field sizes, and beam energy were determined using DoE. The resultant dose distributions obtained using the DoE-selected configuration were compared with the clinical plan.
Results:
The averaged differences between the DoE and clinical plan for 17 patients of doses to 50% of the planning target volume (PTV-D50%), Brain-D60%, Brain-D30%, Brain stem-D1%, Left eye-D1%, Right eye-D1%, Optic nerve-D1%, and Chiasm-D1% were 0.2 ± 0.5%, -1.0 ± 4.6%, 1.7 ± 3.5%, -2.5 ± 6.7%, -0.2 ± 4.9%, -1.2 ± 3.6%, -2.8 ± 7.3%, and -2.1 ± 5.7%, respectively.
Conclusions:
Our optimization workflow obtained using DoE for various field configurations provided the same or slightly superior plan quality compared with that created by experts. This process is feasible for clinical practice and will efficiently improve treatment quality while removing the influence of the planner's experience.
More Related Videos
Related Concept Videos
Arc Length of a Curve
Arc Length Function
Controller Configurations
Control-system compensation involves various configurations, most commonly series or cascade compensation, in which the controller...
Electron Configurations
The relative energies of the subshells determine the order in which atomic orbitals are filled (1s, 2s, 2p, 3s, 3p,...
Electron Configuration of Multielectron Atoms
Arc Length of a Curve: Problem Solving