Biomechanical Properties of Nitinol Staples: Effects of Troughing, Effective Leg Length, and 2-Staple Constructs
R Randall McKnight1, Steve K Lee2, R Glenn Gaston3
1Department of Orthopaedic Surgery, Carolinas Medical Center, Charlotte, NC.
Purpose:
Nitinol memory compression staples are a recent addition to carpal bone fixation. Compared with traditional staples, they have been shown to have superior compression at the far cortex relative to standard and traditional compression staples. The purpose of this study was to (1) determine the effective leg length of different nitinol staples, (2) confirm the effect of 1 versus 2 staples on biomechanical compression, and (3) determine the effect of troughing (countersinking the staple into bone) the bone on staple biomechanical properties.
Methods:
Three commonly used nitinol staples of various bridge and leg lengths were used in a bicortical sawbones block construct. There were 3 separate constructs tested, which included single staple, double staple, and troughed. We measured compression force, stiffness, and bending strength for each construct before and after cyclical 4-point bending. Compression mapping was used to determine the effective leg length of each staple, which included the distance that compression extended beyond the tips of the staple legs.
Results:
Effective leg length for each staple construct extended 2 mm distal to the tip of the shortest staple leg. Two staple constructs more than doubled compressive force and increased bending strength by greater than 90% in all staple types. There was no loss of compressive force before or after loading for single, double, or troughed constructs with any staple type.
Conclusions:
This study supports that nitinol staples do not have to be placed bicortically to achieve adequate compression; placing staples 2 mm short of the far cortex has the same compression as bicortical placement; and troughing of the bone will not significantly diminish the biomechanical properties of the construct.
Clinical Relevance:
Better understanding of the effective leg length of nitinol staples provides support that bicortical placement is not necessary for adequate compression. This study supports troughing bone to minimize implant prominence.
More Related Videos
10:03Uncoupling Coriolis Force and Rotating Buoyancy Effects on Full-Field Heat Transfer Properties of a Rotating Channel
Published on: October 5, 2018
11:47Using 2-Photon Microscopy to Quantify the Effects of Chronic Unilateral Ureteral Obstruction on Glomerular Processes
Published on: March 4, 2022
Related Concept Videos
Buffer Effectiveness
The buffer capacity is the amount of acid or base that can be added to a given volume...
Framing Effects
Biological Effects of Radiation
Effects of feedback
Feedback significantly modifies the gain of a control system. The gain of a system without feedback is altered by a factor of one plus GH, where G represents...
Self-Awareness and Its Effects
Solvating Effects