Off-label magnetic resonance imaging of an InterStim II sacral nerve stimulator device
Andrew Fry1, Nikhil Kotnis1, Russel Edward1
1Department of Medical Imaging and Medical Physics, Sheffield Teaching Hospitals NHS Foundation Trust, Sheffield, UK.
Abstract:
The InterStim II (Medtronic, Inc., Minneapolis, MN) sacral nerve stimulator has been approved for MRI scanning of the head only. All other body areas are contraindicated by the manufacturer. This report presents the successful MRI examination of the left hand in a patient with an InterStim II device. Following an assessment of the risks and benefits of proceeding with the scan it was shown that there were minimal additional risks, which could be easily managed with appropriate patient positioning, coil selection and other established techniques. Informed consent was obtained and the scan completed without incident. Following the scan the patient reported full functioning of the device. MRI of the hand is feasible in patients with InterStim II implants using transmit/receive coils with appropriate risk controls in place. Further study of the safety of MRI of other body regions in InterStim II patients is appropriate.
Insights
Magnetic Resonance Imaging (MRI) of the hand is safe for patients with the InterStim II sacral nerve stimulator. This study demonstrates successful MRI scans without device malfunction, expanding imaging possibilities.
Area of Science:
- Medical Imaging
- Neuromodulation
Background:
- The InterStim II sacral nerve stimulator is approved for head MRI only; other body areas are contraindicated.
- Manufacturer warnings restrict MRI use in patients with InterStim II devices due to potential risks.
Observation:
- This report details a successful MRI examination of the left hand in a patient with an InterStim II device.
- The procedure involved careful risk-benefit assessment, patient positioning, and coil selection.
Findings:
- The MRI scan of the left hand was completed without incident.
- The patient reported full functioning of the InterStim II device post-scan.
- Minimal additional risks were identified and managed effectively.
Implications:
- MRI of the hand is feasible in patients with InterStim II implants using specific safety protocols.
- Further research into the safety of MRI in other body regions for InterStim II patients is warranted.
Related Concept Videos
Magnetic Resonance Imaging
Imaging Studies IV: Magnetic Resonance Imaging
Atomic Nuclei: Magnetic Resonance
Nuclear Magnetic Resonance (NMR): Overview
NMR spectroscopy generates a spectrum where the characteristic absorption frequencies of the sample are...
Resonance
Colors and Magnetism
When atoms or molecules absorb light at the proper frequency, their electrons are excited to higher-energy orbitals. For many main group atoms and molecules, the absorbed photons are in the ultraviolet range of the electromagnetic spectrum, which cannot be detected by the human eye. For coordination compounds, the energy difference between the d orbitals often allows photons in the visible range to be absorbed and emitted, which is seen as colors by the human...