[Continuous Non-Invasive Blood Pressure Measurement Using Pulse Transit Time in Children and Adolescents]
Elena Zagrada1, Tobias Krickau1, Wolfgang Rascher1
1Friedrich-Alexander Universität Erlangen-Nürnberg (FAU), Pediatrics, Erlangen, Germany.
Background:
As an alternative to ambulatory 24-h blood pressure monitors, the continuous, non-invasive blood pressure recording using pulse transit time (PTT) was compared with both oscillometric and intra-arterial measurement.
Methods:
In 144 pediatric patients (9.4±5.2 years) acute blood pressure was determined using three different methods of measurement. In 57 patients (11.4±4.9 years) of the pediatric normal ward, the blood pressure was simultaneously determined continuously by means of PTT compared to the intermittent oscillometric long-term measurement. In 9 patients (9.8±6.8 years) of the pediatric intensive care unit with continuous intra-arterial blood pressure measurement, the blood pressure was measured in parallel by means of PTT.
Results:
Compared to the gold standard sphygmomanometry, mean systolic blood pressure values were comparable to 2 different oscillometric devices, but in the case of diastolic pressure measurement the pressure values were significantly lower or higher than with sphygmomanometry and auscultation. The results of the non-invasive comparative long-term measurements showed better correlations for systolic rather than diastolic pressure. Similar results were obtained comparing the direct intra-arterial blood pressure measurement with the SOMNOtouch™ NIBP device which measures blood pressure using the PTT.
Conclusion:
Continuous, indirect blood pressure recording using pulse transit time provides good correlations to the reference measurements for long-term measurement of blood pressure in children and adolescents and appears to be a usable method to measure the blood pressure of children especially at rest.
Related Concept Videos
Measurement of Blood Pressure
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Blood Pressure
Blood Pressure
The average BP in an adult is typically around 120/80 mmHg (millimeters of mercury). In this measurement, the numerator (120) indicates the systolic pressure, which is the pressure in the arteries during the contraction of the heart's ventricles as blood is expelled. The denominator (80) represents the...
Phase Transitions