Size-Specific Dose Estimation In the Korean Lung Cancer Screening Project: Does a 32-cm Diameter Phantom Represent a
Eun Young Kim1, Tae Jung Kim2, Jin Mo Goo3
1Department of Radiology, Gachon University Gil Medical Center, Incheon 21565, Korea.
Objective:
The purposes of this study were to evaluate size-specific dose estimate (SSDE) of low-dose CT (LDCT) in the Korean Lung Cancer Screening (K-LUCAS) project and to determine whether CT protocols from Western countries are appropriate for lung cancer screening in Korea.
Materials And Methods:
For participants (n = 256, four institutions) of K-LUCAS pilot study, volume CT dose index (CTDIvol) using a 32-cm diameter reference phantom was compared with SSDE, which was recalculated from CTDIvol using size-dependent conversion factor (f-size) based on the body size, as described in the American Association of Physicists in Medicine Report 204. This comparison was subsequently assessed by body mass index (BMI) levels (underweight/normal vs. overweight/obese), and automatic exposure control (AEC) adaptation (yes/no).
Results:
Size-specific dose estimate was higher than CTDIvol (2.22 ± 0.75 mGy vs. 1.67 ± 0.60 mGy, p < 0.001), since the f-size was larger than 1.0 for all participants. The ratio of SSDE to CTDIvol was higher in lower BMI groups; 1.26, 1.37, 1.43, and 1.53 in the obese (n = 103), overweight (n = 70), normal (n = 75), and underweight (n = 4), respectively. The ratio of SSDE to CTDIvol was greater in standard-sized participants than in large-sized participants independent of AEC adaptation; with AEC, SSDE/CTDIvol in large- vs. standard-sized participants: 1.30 ± 0.08 vs. 1.44 ± 0.08 (p < 0.001) and without AEC, 1.32 ± 0.08 vs. 1.42 ± 0.06 (p < 0.001).
Conclusion:
Volume CT dose index based on a reference phantom underestimates radiation exposure of LDCT in standard-sized Korean participants. The optimal radiation dose limit needs to be verified for standard-sized Korean participants.
Related Concept Videos
Estimating Population Standard Deviation
Dose Size and Dosing Frequency: Determination Methods
Estimating Population Mean with Known Standard Deviation
The confidence interval estimate will have the form as follows:
(point estimate - error bound, point estimate +...
Estimating Population Mean with Unknown Standard Deviation
William S. Gosset (1876–1937) of the...
Cell Size
Surface Area
Cells can take in nutrients and water via diffusion through the plasma membrane itself or through specific channels in the membrane. The area of the membrane surrounding...
Brick Sizes
Modular bricks are the most common type and are sized to include the mortar joint, which is essential for...