Titanium Fastener Utilization During HeartMate 3 Left Ventricular Assist Device Implantation
Carl A Johnson1, Amber L Melvin, Davida A Robinson
1From the Division of Cardiac Surgery, University of Rochester Medical Center, Rochester, NY USA.
Titanium fasteners offer a feasible and efficient method for securing the inflow sewing ring during HeartMate 3 implantation, leading to reduced cardiopulmonary bypass times compared to traditional hand-tied knots.
Area of Science:
- Cardiovascular Surgery
- Medical Device Technology
- Biomaterials Engineering
Background:
- The HeartMate 3 is a left ventricular assist device (LVAD) requiring secure attachment of its inflow sewing ring.
- Traditional methods for securing the sewing ring involve hand-tied knots, which can be time-consuming.
- Investigating novel, efficient methods for device securement is crucial for improving surgical outcomes.
Purpose of the Study:
- To assess the feasibility of using titanium fasteners for HeartMate 3 inflow sewing ring securement.
- To compare surgical times, specifically cardiopulmonary bypass and total operative durations, between titanium fasteners and hand-tied knots.
Main Methods:
- In vitro assessment of clearance between the sewing ring and HeartMate 3 device for titanium fasteners.
- Retrospective review of 31 patients undergoing HeartMate 3 implantation, divided into hand-tied knot (n=18) and titanium fastener (n=13) groups.
- Standardized cannulation and implantation techniques, with variations only in sewing ring securement method.
Main Results:
- Adequate in vitro clearance was confirmed for titanium fastener use.
- Successful in vivo securement of the inflow sewing ring was achieved with titanium fasteners.
- Cardiopulmonary bypass time was significantly shorter in the titanium fastener group (75 min) compared to the hand-tied group (92 min; P < 0.03).
- Total operative time showed no significant difference between the groups (177 min vs. 193 min; P = 0.513).
Conclusions:
- Titanium fasteners provide an efficient and effective method for securing the HeartMate 3 inflow sewing ring.
- The use of titanium fasteners is associated with reduced cardiopulmonary bypass times.
- Titanium fasteners do not significantly alter total operative time compared to hand-tied knots.
Related Concept Videos
Wood Fasteners
Nails are the most common fasteners, consisting of sharp-pointed metal pins that are driven into wood using a hammer or a mechanical nail gun. They come in a variety of diameters, heads, and lengths. Nailing techniques include...
Steel Fastening Techniques
Rivets are cylindrical steel fasteners with a specially designed head. During application, rivets are heated until white-hot and then inserted through pre-drilled holes in the steel sections. A pneumatic hammer is used to shape the exposed end into a second head, securing the sections together.
Bolting is another...
ortho–para-Directing Activators: –CH3, –OH, –⁠NH2, –OCH3
Non-ohmic Devices
Consider a simple circuit consisting of a battery, a diode, and a resistor. A...
Mnemonic Devices
Acronyms
Acronyms are created by using the initial letters of a series of words to form a new word or phrase. This approach condenses complex information into a single, memorable entity. For example,...
Piaget's Stage 3 of Cognitive Development
Conservation and Constancy of Quantity
A significant cognitive milestone in the...