Towards non-invasive characterisation of coronary stent re-endothelialisation - An in-vitro, electrical impedance
Ian Holland1, Christopher McCormick1, Patricia Connolly1
1Department of Biomedical Engineering, University of Strathclyde, Glasgow, Scotland, United Kingdom.
Abstract:
The permanent implantation of a stent has become the most common method for ameliorating coronary artery narrowing arising from atherosclerosis. Following the procedure, optimal arterial wall healing is characterised by the complete regrowth of an Endothelial Cell monolayer over the exposed stent surface and surrounding tissue, thereby reducing the risk of thrombosis. However, excessive proliferation of Smooth Muscle Cells, within the artery wall can lead to unwanted renarrowing of the vessel lumen. Current imaging techniques are unable to adequately identify re-endothelialisation, and it has previously been reported that the stent itself could be used as an electrode in combination with electrical impedance spectroscopic techniques to monitor the post-stenting recovery phase. The utility of such a device will be determined by its ability to characterise between vascular cell types. Here we present in-vitro impedance spectroscopy measurements of pulmonary artery porcine Endothelial Cells, Human Umbilical Vein Endothelial Cells and coronary artery porcine Smooth Muscle Cells grown to confluence over platinum black electrodes in clinically relevant populations. These measurements were obtained, using a bespoke impedance spectroscopy system that autonomously performed impedance sweeps in the 1kHz to 100kHz frequency range. Analysis of the reactance component of impedance revealed distinct frequency dependent profiles for each cell type with post confluence reactance declines in Endothelial Cell populations that have not been previously reported. Such profiles provide a means of non-invasively characterising between the cell types and give an indication that impedance spectroscopic techniques may enable the non-invasive characterisation of the arterial response to stent placement.
Related Concept Videos
Acute Coronary Syndrome III: Diagnostic Studies
Impedance Combination
Impedances and Admittance
Impedance is a measure of resistance to sinusoidal current flow in an AC circuit. Unlike their behavior in DC circuits, where inductors appear as short circuits and capacitors as open circuits, the behavior of these components in AC...
Coronary Circulation
Coronary circulation begins at the base of the aorta, where two main arteries arise—the left and right coronary arteries. These arteries encircle the heart in the coronary sulcus and supply the...
Series Impedances: Three-Phase Line
Using Kirchhoff's laws, an integro-differential equation for the network is derived. This equation accounts for unbalanced phase currents, which may induce return currents through neutral wires and the earth, seeking the least impedance path. Earth return conductors can replace the...
Bus Impedance Matrix
In the first circuit, all machine voltage sources are short-circuited, leaving only the prefault voltage source at the fault location. The positive-sequence bus impedance matrix can be determined by solving the nodal equations,...