Measuring Diffusion Coefficients via Two-photon Fluorescence Recovery After Photobleaching
Published on: February 26, 2010
Apparent diffusion coefficient reproducibility in brain tumors measured on 1.5 and 3 T clinical scanners: A pilot
Zulejha Merhemic1, Bilal Imsirovic2, Nurija Bilalovic3
1University of Sarajevo, Department of Radiology, Kranjceviceva 12, 71000 Sarajevo, Bosnia and Herzegovina.
Background:
Gradient and coil systems, pulse sequence design, and imaging parameters, as well as different scanners, can influence apparent diffusion coefficient (ADC) values. The aim of this study was to evaluate the effect of two different field strengths on the reproducibility of mean absolute ADC measurements in various primary and secondary brain tumors.
Methods:
Fifty patients with histologically proven brain tumors were prospectively examined on two MR scanners from the same vendor, with different field strengths-1.5T and 3T-on the same day. Absolute ADC values were compared using the Wilcoxon matched-pairs signed-rank test. Inter-scanner agreement between two different fields in the same tumor was examined using correlation coefficients, and the discrepancy between the highest and the lowest mean absolute ADC values between scanners was tested using a one-way analysis of variance. Statistical significance was set at p < 0.05.
Results:
There was no statistically significant difference between mean absolute ADC values obtained on 1.5T and 3T scanners for all patients and all brain tumor types. The intratumoral difference in ADC values, averaged from two scanners in the same tumor type, ranged from 1.58 to 4.5% for 1.5T, and from 1.18 to 4.37% for 3T.Inter-scanner agreement was high, and the kappa coefficient ranged from 0.88 to 0.99, with no significant difference between obtained values on different field strengths.
Conclusion:
Based on the results obtained in our study, there is no significant difference between mean absolute ADC values measured in various primary and secondary brain tumors at different field strengths (1.5 and 3.0T MR systems), in the same patient, and in the same tumor, measured on the same day.
Related Concept Videos
Apparent Weight
Consider a person standing on a bathroom scale inside an elevator. If the scale is accurate at rest, its reading equals the...
Diffusion
Pilot and Numeric Relaying
Coefficient of Variation
The coefficient of variation is a practical statistical tool in finance. It allows investors to assess the volatility or...
Coefficient of Correlation
If you suspect a linear relationship between x and y, then r can measure how strong the linear relationship is.
What the VALUE of r tells us:
The value of r is always between –1 and +1: –1 ≤ r ≤ 1.
The size of the correlation r indicates the...
Confidence Coefficient