Initial operation results of NE213 scintillation detector for time-resolved measurements on triton burnup in KSTAR
Jungmin Jo1, MunSeong Cheon2, Junghee Kim2
1Department of Energy Systems Engineering, Seoul National University, Seoul 151-744, South Korea.
Abstract:
In time-resolved measurement for triton burnup in Korea Superconducting Tokamak Advanced Research (KSTAR) deuterium plasmas, an NE213 liquid scintillation detector was installed and operated during the 2017 KSTAR campaign. The detector is composed of an NE213 scintillator (50 mm in diameter and 10 mm in thickness) and a photomultiplier tube (PMT). The PMT anode signal was processed under a data acquisition system which contains a field programmable gate array circuit and pulse processing software that is capable of discriminating gamma-ray and neutron pulse signals. In order to determine an appropriate threshold level for the 14 MeV neutron signal resulting from triton burnup, the NE213 scintillation detector was calibrated by using d-d and d-t neutron generators at the National Fusion Research Institute and Intense 14 MeV Neutron Source Facility, OKTAVIAN, Osaka University, Japan. The detector was installed on KSTAR with a 10 mm thick soft-iron stray magnetic field shield and a radiation shield which consists of 100 mm thick lead blocks and 200 mm thick borated polyethylene blocks. A discrimination range for d-t neutron was determined based on test results from neutron generators and KSTAR. Data points selected from the discrimination range were consistent with the classical triton confinement characteristics. In conclusion, under condition of an input counting rate of 1.9 × 105 counts per second (CPS), the detector is able to measure triton burnup signals up to 500 CPS for various plasma parameters.
Related Concept Videos
Initiation of Translation
First, the initiator tRNA must be selected from the pool of elongator tRNAs by eukaryotic initiation factor 2 (eIF2). The initiator tRNA (Met-tRNAi) has conserved sequence elements including modified bases at...
Initiation of Translation
Gas Chromatography: Types of Detectors-I
TCD is the earliest and most widely used detector that operates by measuring the changes in the thermal conductivity of the carrier gas. When a sample compound enters the detector,...
Transcription Initiation
The promoters and enhancers and their accessory proteins allow tight regulation of...
Gas Chromatography: Overview of Detectors
A non-destructive detector allows a sample to be analyzed without altering or consuming it, meaning the sample can be collected after detection for further analysis. Examples include thermal conductivity detectors and...
Gas Chromatography: Types of Detectors-II