Upgrades to the in-vessel calibration light source on JET
N J Conway1, A J Cackett1, C F Maggi1
1UKAEA/CCFE, Culham Science Centre, Abingdon, Oxon OX14 3DB, United Kingdom.
The Joint European Torus (JET) upgraded its in-vessel calibration light source (ICLS) for improved diagnostic calibration. Enhancements include a shorter cable, on-board spectrometer, and internal baffles for better spectral uniformity.
Area of Science:
- Fusion Energy Research
- Plasma Diagnostics
- Optical Engineering
Background:
- The Joint European Torus (JET) utilizes an in-vessel calibration light source (ICLS) for diagnostic calibration across UV, visible, and IR wavelengths.
- The ICLS is an integrating sphere positioned within the vacuum vessel using a remote handling (RH) system during shutdowns.
Purpose of the Study:
- To enhance the efficiency, quality, and operational flexibility of the ICLS system at JET.
- To improve spectral radiance uniformity and enable real-time verification of calibration light source performance.
Main Methods:
- Replaced a 20m umbilical cable with a 2m lightweight cable for direct connection to the RH manipulator.
- Integrated an on-board micro-spectrometer for real-time spectral verification.
- Installed new baffles within the integrating sphere to improve spectral radiance uniformity.
Main Results:
- Reduced calibration deployment time and improved operational flexibility by eliminating the need for vessel access floor installation.
- Enabled compatibility with calibrations following high neutron-fluence operations.
- Significantly improved spectral radiance uniformity at non-normal viewing angles, compensating for RH orientation uncertainties.
Conclusions:
- The upgraded ICLS system at JET demonstrates enhanced performance and operational efficiency.
- The modifications ensure reliable and accurate diagnostic calibration, even after intense operational periods.
- The improved uniformity and real-time verification contribute to more robust fusion energy research.
Related Concept Videos
Free Jet
Atomic Absorption Spectroscopy: Radiation and Light Sources
Two common narrow-range 'line' sources used in AAS are hollow-cathode lamps (HCLs) and...
Glassware Calibration
Volumetric flasks: Volumetric flasks are designed to prepare aqueous solutions of precise volumes accurately with a calibration line on the neck. To calibrate a volumetric flask, it is important to fill it with distilled...
Instrument Calibration
Analytical Balance Calibration
An analytical balance measures mass and requires regular calibration to...
Calibration Curves: Correlation Coefficient
Plotting and Calibrating the Root Locus
The maximum gain occurs at the breakaway points between open-loop poles on the real axis, while the minimum gain is...