The behavior of polyetheretherketone healing abutments when measuring implant stability with electronic percussive
Çağatay Dayan1, Canan Bural2, Onur Geçkili2
1Program of Dental Technicians, Istanbul Cerrahpasa University, Istanbul, Turkey.
Background:
It is unknown whether it is possible to measure implant stability with polyetheretherketone (PEEK) healing abutments using electronic percussive testing (EPT).
Purpose:
To investigate the reliability of the percussive test values (PTVs) measured with PEEK healing abutments and to compare them with the PTVs measured with titanium healing abutments.
Material And Methods:
Thirty dental implants were inserted into the fresh pelvis belonging to a cow. Titanium healing abutments (2 and 5 mm), PEEK healing abutments (5 mm), and prepable standard titanium abutments (5 mm) were screwed to the implants, respectively, and PTVs were measured by two examiners using a wireless EPT device. Differences in PTVs between different dental implant components were evaluated using Friedman's test with post hoc Wilcoxon signed-rank test and Bonferroni correction. Inter and intra-observer reliabilities were detected using interclass correlation coefficients (ICCs) RESULTS: The mean PTVs obtained using the PEEK healing abutments were significantly higher than the mean PTVs obtained using the other abutments for both examiners (P < .01). The ICCs for intra-observer reliability were detected as poor for PEEK healing abutments; and excellent for the other abutment types for both examiners. The ICCs for the inter-observer reliability between the two examiners were poor (0.25) for PEEK healing abutments, and excellent for the 2 and 5 mm titanium healing abutments and for standard abutments (0.82, 0.84 and 0.93, respectively).
Conclusion:
Within the limitations of this in vitro study, it may be concluded that EPT measurements should be avoided using PEEK healing abutments because of the poor reliability.
Related Concept Videos
Electron Behavior
Electrons are negatively charged subatomic particles that are attracted to an orbit around the positively-charged nucleus of an atom. They reside in locations that are associated with energy levels called shells and are further organized into sub-shells and orbitals within each shell.
Electrons Orbit the Nucleus
Electrons are found in specific locations outside of the nucleus. The shell in which an electron resides indicates the general energy level of the electron: those closer to the...
Electron Behavior
Electrons Orbit the Nucleus
Electrons are found in specific locations outside of the nucleus. The shell in which an electron resides indicates the general energy level of the electron: those closer to the nucleus have less energy,...
Nuclear Stability
To hold positively charged protons together...
Assessment of the Abdomen II: Percussion
Percussion
Percussion is an essential...
Combustion Energy: A Measure of Stability in Alkanes and Cycloalkanes
Alkanes undergo combustion in the presence of excess oxygen and high-temperature conditions to give carbon dioxide and water. A combustion reaction is the energy source in natural gas, liquified...
Electronic Distance Measuring Instruments