Characterization of Recombination Effects in a Liquid Ionization Chamber Used for the Dosimetry of a Radiosurgical Accelerator
Published on: May 9, 2014
The characterization of a large multi-axis ionization chamber array in a 1.5 T MRI linac
T J Perik1, J J Kaas1, S Greilich2
1Department of Radiation Oncology, The Netherlands Cancer Institute, Plesmanlaan 121, 1066CX Amsterdam, Netherlands.
None:
By combining magnetic resonance imaging (MRI) scanners and radiotherapy treatment units the need arises for new radiation measurement equipment that can be used in the magnetic field of the MRI. This study describes the investigation of the influence of the 1.5 T magnetic field from an MRI linac on the STARCHECKMAXI MR, a large 2D ionization chamber detector panel. Measurements were performed on an MRI linac and a conventional linac to investigate the behaviour of the detector panel with and without the 1.5 T magnetic field. We measured reproducibility, linearity, warm-up effect, saturation/recombination and chamber orientation. A comparison with gafchromic film was performed and the effect of motion of the panel during measurements inside a magnetic field was investigated. The reproducibility, linearity, warm-up effect, saturation/recombination show no significant deviations with or without magnetic field. An absolute difference in reading of 2.1% was found between off-axis chambers on different axes. The comparison with film shows good agreement. Spurious readings are induced while the panel is undergoing a motion in the magnetic field during measurements. The STARCHECKMAXI MR is suited for use in a 1.5 T MRI linac. Care must be taken when comparing un-normalized profiles from different axes of the detector panel and when the panel is undergoing motion during measurements.
Related Concept Videos
Ionization Energy
Hypothalamic-Pituitary Axis
Perpendicular-Axis Theorem
Consider a circular disc of mass M and radius R lying along an x-y plane. The origin lies at the center of the disc, and the z-axis is perpendicular to the disc's plane. All three axes coincide at the disc's center. The moment of inertia of this...
Parallel-axis Theorem
Chambers of the Heart
Deoxygenated blood from the body is received in the right...
Load along a Single Axis
Consider a beam of length L subjected to a varying load, which is a combination of parabolic and trapezoidal load distribution along the x-axis. In this case, it is essential to determine the resultant loads, their locations, and...