Intersubject specific absorption rate variability analysis through construction of 23 realistic body models for
Ettore F Meliadò1,2, Cornelis A T van den Berg1, Peter R Luijten1
1Center for Image Sciences, University Medical Center Utrecht, Utrecht, The Netherlands.
Purpose:
For ultrahigh field (UHF) MRI, the expected local specific absorption rate (SAR) distribution is usually calculated by numerical simulations using a limited number of generic body models and adding a safety margin to take into account intersubject variability. Assessment of this variability with a large model database would be desirable. In this study, a procedure to create such a database with accurate subject-specific models is presented. Using 23 models, intersubject variability is investigated for prostate imaging at 7T with an 8-channel fractionated dipole antenna array with 16 receive loops.
Method:
From Dixon images of a volunteer acquired at 1.5T with a mockup array in place, an accurate dielectric model is built. Following this procedure, 23 subject-specific models for local SAR assessment at 7T were created enabling an extensive analysis of the intersubject B1+ and peak local SAR variability.
Results:
For the investigated setup, the maximum possible peak local SAR ranges from 2.6 to 4.6 W/kg for 8 × 1 W input power. The expected peak local SAR values represent a Gaussian distribution W/kg) with realistic prostate-shimmed phase settings and a gamma distribution Γ(24,0.09) with multidimensional radiofrequency pulses. Prostate-shimmed phase settings are similar for all models. Using 1 generic phase setting, average B1+ reduction is 7%. Using only 1 model, the required safety margin for intersubject variability is 1.6 to 1.8.
Conclusion:
The presented procedure allows for the creation of a customized model database. The results provide valuable insights into B1+ and local SAR variability. Recommended power thresholds per channel are 3.1 W with phase shimming on prostate or 2.6 W for multidimensional pulses.
More Related Videos
Related Concept Videos
Metabolic States of the Body: The Absorptive State
Variability: Analysis
The range is a simple measure of variability, indicating the difference between the highest and...
One-Compartment Open Model for Extravascular Administration: Zero-Order Absorption Model
Zero-order absorption maintains a steady rate irrespective of the amount of drug left to be absorbed, making it a constant process. In the...
One-Compartment Open Model for Extravascular Administration: First-Order Absorption Model
Noncompartmental Analysis: Mean Transit, Absorption and Dissolution Time
One of the key parameters is the mean transit time (MTT), which refers to the total duration required for drug molecules to transit through the body. MTT is determined by calculating the ratio of the area under the moment curve to the area...
Reaction Rate
The mathematical representation of the change in the concentration of reactants and products, over time, is the rate...