Accelerated, first-pass cardiac perfusion pulse sequence with radial k-space sampling, compressed sensing, and
Nivedita K Naresh1, Hassan Haji-Valizadeh1,2, Pascale J Aouad1
1Department of Radiology, Northwestern University Feinberg School of Medicine, Chicago, Illinois.
Purpose:
To develop an accelerated cardiac perfusion pulse sequence and test whether it is capable of increasing spatial coverage, generating high-quality images, and enabling quantification of myocardial blood flow (MBF).
Methods:
We implemented an accelerated first-pass cardiac perfusion pulse sequence by combining radial k-space sampling, compressed sensing (CS), and k-space weighted image contrast (KWIC) filtering. The proposed and clinical standard pulse sequences were evaluated in a randomized order in 13 patients at rest. For visual analysis, 3 readers graded the conspicuity of wall enhancement, artifact, and noise level on a 5-point Likert scale (overall score index = sum of 3 individual scores). Resting MBF was calculated using a Fermi function model with and without KWIC filtering. Mean visual scores and MBF values were compared between sequences using appropriate statistical tests.
Results:
The proposed pulse sequence produced greater spatial coverage (6-8 slices) with higher spatial resolution (1.6 × 1.6 × 8 mm3 ) and shorter readout duration (78 ms) compared to clinical standard (3-4 slices, 3 × 3 × 8 mm3 , 128 ms, respectively). The overall image score index between accelerated (11.1 ± 1.3) and clinical standard (11.2 ± 1.3) was not significantly different (P = 0.64). Mean resting MBF values with KWIC filtering (0.9-1.2 mL/g/min across different slices) were significantly lower (P < 0.0001) than those without KWIC filtering (3.1-4.3 mL/g/min) and agreed better with values reported in literature.
Conclusion:
An accelerated, first-pass cardiac perfusion pulse sequence with radial k-space sampling, CS, and KWIC filtering is capable of increasing spatial coverage, generating high-quality images, and enabling quantification of MBF.
More Related Videos
Related Concept Videos
Space Trusses
At the core of a space truss lies the fundamental unit known as the tetrahedron. This structure is composed of six members that form a three-dimensional shape...
State Space Representation
Consider an RLC circuit, a...
Assessment of radial pulse
The radial pulse, located at the wrist, is often the preferred site for assessing peripheral pulse because of its accessibility and dependability. The process of determining the radial pulse involves several steps:
Space Trusses: Problem Solving
Consider a tripod consisting of a tetrahedral space truss with a ball-and-socket joint at C. Suppose the height and lengths of the horizontal and vertical...
Transfer Function to State Space
In an RLC...
State Space to Transfer Function
The transformation process begins with the state-space representation, characterized by the state equation and the output equation. These equations are typically represented as: