The Quantification of Injectability by Mechanical Testing
Published on: May 13, 2020
Quantification of injection force mechanics during injection laryngoplasty
Joseph J Pearson1, Alexandra S Ortiz2, Sergio Montelongo1
1Department of Biomedical Engineering, University of Texas at San Antonio, San Antonio.
Objectives:
In-office or operative injection laryngoplasty requires needle stability for accurate material placement. To date, no reports compare injection forces based on needle gauge, bends, length, or material type or temperature. We hypothesize these factors alter injection forces and could impact clinical use.
Methods:
Swine larynges were placed in a compression testing machine. Syringes were affixed to a stabilizing crossbeam. Straight needles (25G 1.5-inch; 27G 1.25-inch; or 9.8-inch malleable shaft 16G per oral with 24G tapered needle tip) were inserted into the swine vocal folds to simulate realistic tissue resistance pressure. Compressive loading was conducted at 40 mm/minute until steady-state force was achieved. Tests were completed with calcium hydroxylapatite (CaHa), carboxymethylcellulose, and hyaluronic acid at various temperatures and CaHa with various bends in the needles (n = 3 per group, comparisons performed by two-way analysis of variance (ANOVA), Tukey's post-hoc).
Results:
Needle size, shape, and temperature altered injection force. Steady-state force was highest with the per-oral needle at a mean of 44.55N compared to 26.44N and 29.77N in the 25G and 27G percutaneous needles, respectively (P < 0.001). Stiffness rate (initial increasing force vs. distance to initiate injection) ranged from 19.75N/mm (per oral) to 22.06N/mm (25G) to 24.56N/mm (27G), (P = 0.875). Adding multiple bends to the per-oral needle increased stiffness rate to 24.99N/mm (P = 0.035), whereas the 25G needle stiffness rate remained unchanged (P = 0.941), with the stiffness rate decreasing in the 27G needle with increasing bends (P = 0.033). Increased temperature decreased injection forces across all materials.
Conclusion:
Needle caliber, length, and bends impact steady-state forces and stiffness rates during vocal fold injection.
Level Of Evidence:
NA Laryngoscope, 129:1060-1066, 2019.
Related Concept Videos
Cell-matrix's Response to Mechanical Forces
Anchoring junctions mechanically attach a cell to the...
Gas Chromatography: Sample Injection Systems
Two primary injection methods are used...
Nonlinear Pharmacokinetics: Drug Elimination for IV Bolus Injection
Following the administration of a single intravenous (IV) bolus injection, we can determine the concentration of the drug in the plasma at any given time. This calculation is achieved using a specific equation that integrates the values of Vmax and KM.
We can also...
Drug Accumulation During Multiple Dosing: Repetitive IV Injections
Intermolecular Forces
Electromotive Force