QRS Complex Detection and Measurement Algorithms for Multichannel ECGs in Cardiac Resynchronization Therapy Patients
Antonia E Curtin1, Kevin V Burns2, Alan J Bank2
1Biomedical Engineering DepartmentUniversity of MinnesotaMinneapolisMN55455USA.
Abstract:
We developed an automated approach for QRS complex detection and QRS duration (QRSd) measurement that can effectively analyze multichannel electrocardiograms (MECGs) acquired during abnormal conduction and pacing in heart failure and cardiac resynchronization therapy (CRT) patients to enable the use of MECGs to characterize cardiac activation in such patients. The algorithms use MECGs acquired with a custom 53-electrode investigational body surface mapping system and were validated using previously collected data from 58 CRT patients. An expert cohort analyzed the same data to determine algorithm accuracy and error. The algorithms: 1) detect QRS complexes; 2) identify complexes of the most prevalent morphology and morphologic outliers; and 3) determine the array-specific (i.e., anterior and posterior) and global QRS complex onsets, offsets, and durations for the detected complexes. The QRS complex detection algorithm had a positive predictivity and sensitivity of ≥96% for complex detection and classification. The absolute QRSd error was 17 ± 14 ms, or 12%, for array-specific QRSd and 12 ± 10 ms, or 8%, for global QRSd. The absolute global QRSd error (12 ms) was less than the interobserver variation in that measurement (15 ± 10 ms). The sensitivity, positive predictivity, and error of the algorithms were similar to the values reported for current state-of-the-art algorithms designed for and limited to simpler data sets and conduction patterns and within the variation found in clinical 12-lead ECG QRSd measurement techniques. These new algorithms permit accurate, real-time analysis of QRS complex features in MECGs in patients with conduction disorders and/or pacing.
Related Concept Videos
Correlation between ECG and Cardiac Cycle
A cardiac action potential originates in the SA node and spreads throughout the atria and the AV node in approximately 0.03 seconds. This results in the P wave in an ECG and triggers atrial contraction. The action potential is then briefly slowed at the AV node, allowing the atria to contract and fill the ventricles with blood before...
ECG Interpretation of Rhythms
Components of the Electrocardiogram
The primary components of a normal ECG waveform in Normal sinus rhythm(NSR) include the P wave, PR interval, QRS complex, ST segment, T wave, and occasionally a U wave.
ECG waveforms are divided by vertical and horizontal lines at standard intervals.
The horizontal axis measures time and rate, and the vertical axis measures amplitude or voltage....
Trial and Error and Algorithm
Gene Therapy
ECG Interpretation of Arrhythmias I: Sinus Arrhythmias
Types of Arrhythmias
Sinus Node Arrhythmias
Sinus Bradycardia: Originating from the sinoatrial (SA) node, sinus bradycardia involves slower impulses, resulting in a heart rate of less than 60 beats per minute (bpm). Causes include sleep, vagal stimulation, beta-blockers, hypothyroidism,...
Protein Complex Assembly
Many viruses self-assemble into a fully functional unit using the infected host cell to...