Related Experiment Video
Updated: Oct 17, 2026

Patient Directed Recording of a Bipolar Three-Lead Electrocardiogram using a Smartwatch with ECG Function
Published on: December 11, 2019
Evaluation of a Wrist-Worn Device for ECG-Based Detection of Acute Myocardial Infarction
Karolina Eitminaviciene1, Daivaras Sokas1, Zygimantas Abramikas2,3
1Biomedical Engineering InstituteKaunas University of Technology Kaunas 44249 Lithuania.
Objective:
To determine whether a two-lead ECG (wECG), acquired using a self-applicable, wrist-worn device, provides diagnostic value in acute myocardial infarction (AMI) detection.
Methods:
Seventy-one patients with AMI, 26 with other cardiovascular diseases (CVDs), and 54 healthy controls were enrolled in the study. To acquire lead I, one electrode of the wrist-worn device was touched with the index finger of the opposite hand, while another lead was acquired by touching the electrode on the strap to a specific part of the body. The diagnostic value of the wECG is evaluated either directly or through synthesis of the standard 12-lead ECG using an echo state network.
Results:
Using the standard ECG, an AMI detector based on a convolutional neural network yields a sensitivity of 0.86 and a specificity of 0.71; the corresponding figures for blinded cardiologist diagnosis are 0.77/0.88. Using the synthesized ECG, the V3 electrode touch site yields the best performance, with a sensitivity/specificity of 0.84/0.71 for AMI detection and 0.67/0.86 for cardiologist diagnosis. Compared to the synthesized ECG across all touch sites, use of the wECG results in a substantially lower sensitivity, reduced by 0.12-0.13, while the specificity is comparable or slightly higher. Implementation of a two-stage approach, consisting of detector-based screening followed by cardiologist review of positive cases, results in correct decisions in 63% of the AMI cases using either the synthesized ECG or the wECG.
Conclusion:
The two-lead ECG acquired with a wrist-worn device offers diagnostic value and enables screening for AMI.
More Related Videos
18:11A Research Method For Detecting Transient Myocardial Ischemia In Patients With Suspected Acute Coronary Syndrome Using Continuous ST-segment Analysis
Published on: December 28, 2012
04:24A Novel Digital Platform for a Monitored Home-based Cardiac Rehabilitation Program
Published on: April 19, 2019
Related Concept Videos
Electrocardiogram
Three major waveforms are present in a typical ECG recording: the P wave, the QRS complex, and the T...
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac muscle...
Acute Coronary Syndrome III: Diagnostic Studies
Holter Monitor: 24-Hour Monitoring
Dysrhythmias V: Evaluating Dysrhythmias
Assessment of apical radial pulse
The A-R pulse assessment involves simultaneous evaluation of the apical and radial pulses. When the apical and radial pulse rates vary, this assessment helps identify a pulse deficit.
Pre-Procedural Preparation