Related Experiment Video
Updated: Feb 2, 2026

Single Cell Measurement of Dopamine Release with Simultaneous Voltage-clamp and Amperometry
Published on: November 21, 2012
Self-releasing of a titanium clamp (CranioFix) in cranioplasty
Minwook Yoo1, Sungjoon Lee1, Sung-Chul Jin1
1a Department of Neurosurgery , Inje University Haeundae Paik Hospital , Busan City , Republic of Korea.
Abstract:
The CranioFix® titanium clamp was developed for fixing bone flaps after craniotomy. In this case, to prevent cosmetic problems, the bone flaps were trimmed with a beveled edge, and a CranioFix clamp was used for bone flap fixation. As the brain swelled, the CranioFix clamp became loosened. If brain swelling is anticipated and bone margins have been trimmed, it may be better to consider other fixation methods.