A dosimetric analysis of a spine SBRT specific treatment planning system
Daniel L Saenz1, Richard Crownover1, Sotirios Stathakis1
1University of Texas Health San Antonio, Mays Cancer Center, San Antonio, TX, USA.
Purpose:
The Brainlab Elements treatment planning system utilizes distinct modules for treatment planning specific to stereotactic treatment sites including single or multiple brain lesions as well as spine. This work investigates the hypothesis that an optimization tailored specifically to spine can in fact create dosimetrically superior plans to those created in more general use treatment planning systems (TPS).
Methods:
Ten spine patients at our institution were replanned in Brainlab Elements, Phillips Pinnacle3 , and Elekta Monaco. The planning target volume (PTV) included the vertebral body (in either the thoracic or lumbar spine), pedicles, and transverse processes. In all plans, the target was prescribed 20 Gy to 95% of the PTV. Objectives for the study included D5%<25 Gy and spinal cord D0.035cc < 14 Gy. Plans were evaluated by the satisfaction of the objectives as well total monitor units (MU), gradient index (GI), conformity index (CI), and dose gradient (distance between 100% and 50% isodose lines) in a selected slice between the vertebral body and spinal cord.
Results:
All TPS produced clinically acceptable plans. The sharpest dose gradient was achieved with Elements (mean 3.3 ± 0.2 mm). This resulted in lowest spinal cord maximum point doses (6.6 ± 1.0 Gy). Gradient indices were also the smallest for Elements (3.6 ± 0.5). Further improvement in gradient index and spinal cord sparing were not performed due to the subsequent violation of the PTV D5% < 25 Gy constraint or the loss of conformity due to the loss of coverage at the PTV-spinal canal interface.
Conclusions:
Brainlab Elements planning which relies on arc duplication to specifically optimize for spine anatomy did result in dosimetrically superior plans while holding prescription levels constant. While any planning system can improve upon specific dosimetric objectives, the simultaneous satisfaction of all constraints was best achieved with Brainlab Elements.
Related Concept Videos
Sampling Plans
Random sampling is a method where each member of the population has an equal chance of being selected for the sample. It involves selecting individuals randomly, often using random number generators or lottery-type methods. For example, when analyzing the properties of a...
Planning Nursing Care I
Planning Nursing Care II
Specific Heat
For example, increasing the temperature of one gram of water by 1°C requires one calorie of heat energy and can be written as 1 cal/g-°C, or...
Role of Communication in the Nursing Process II: Planning and Implementation
Nursing Process for Patient and Caregiver Teaching II: Planning and Implementation