High-Resolution Chest Computed Tomography Imaging of the Lungs: Impact of 1024 Matrix Reconstruction and
David J Bartlett, Chi Wan Koo, Brian J Bartholmai
1Siemens Healthineers, Malvern, PA.
Objectives:
The aim of this study was to evaluate if a high-resolution photon-counting detector computed tomography (PCD-CT) system with a 1024×1024 matrix reconstruction can improve the visualization of fine structures in the lungs compared with conventional high-resolution CT (HRCT).
Materials And Methods:
Twenty-two adult patients referred for clinical chest HRCT (mean CTDI vol, 13.58 mGy) underwent additional dose-matched PCD-CT (mean volume CT dose index, 13.37 mGy) after written informed consent. Computed tomography images were reconstructed at a slice thickness of 1.5 mm and an image increment of 1 mm with our routine HRCT reconstruction kernels (B46 and Bv49) at 512 and 1024 matrix sizes for conventional energy-integrating detector (EID) CT scans. For PCD-CT, routine B46 kernel and an additional sharp kernel (Q65, unavailable for EID) images were reconstructed at 1024 matrix size. Two thoracic radiologists compared images from EID and PCD-CT noting the highest level bronchus clearly identified in each lobe of the right lung, and rating bronchial wall conspicuity of third- and fourth-order bronchi. Lung nodules were also compared with the B46/EID/512 images using a 5-point Likert scale. Statistical analysis was performed using a Wilcoxon signed rank test with a P < 0.05 considered significant.
Results:
Compared with B46/EID/512, readers detected higher-order bronchi using B46/PCD/1024 and Q65/PCD/1024 images for every lung lobe (P < 0.0015), but in only the right middle lobe for B46/EID/1024 (P = 0.007). Readers were able to better identify bronchial walls of the third- and fourth-order bronchi better using the Q65/PCD/1024 images (mean Likert scores of 1.1 and 1.5), which was significantly higher compared with B46/EID/1024 or B46/PCD/1024 images (mean difference, 0.8; P < 0.0001). The Q65/PCD/1024 images had a mean nodule score of 1 ± 1.3 for reader 1, and -0.1 (0.9) for reader 2, with one reader having improved nodule evaluation scores for both PCD kernels (P < 0.001), and the other reader not identifying any increased advantage over B46/EID/1024 (P = 1.0).
Conclusions:
High-resolution lung PCD-CT with 1024 image matrix reconstruction increased radiologists' ability to visualize higher-order bronchi and bronchial walls without compromising nodule evaluation compared with current chest CT, creating an opportunity for radiologists to better evaluate airway pathology.
More Related Videos
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies III: Computed Tomography
Electron Microscope Tomography and Single-particle Reconstruction
Electron Tomography
Electron tomography can be performed either in TEM or STEM (scanning transmission...
Design Example: Traverse Angle Computations
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
Area Computation by the Alternative Coordinate Method