Improvement of Detection Sensitivity of Microbubbles as Sensors to Detect Ambient Pressure
Fei Li1,2, Deyu Li3, Fei Yan4,5
1Paul C. Lauterbur Research Center for Biomedical Imaging, Shenzhen Institutes of Advanced Technology, Chinese Academy of Sciences, Shenzhen 518055, China. fei.li@siat.ac.cn.
Abstract:
Microbubbles are considered a promising tool for noninvasive estimation of local blood pressure. It is reported that the subharmonic scattering amplitude of microbubbles decreases by 9 to 12 dB when immersed in the media under an ambient pressure variation from 0 to 180 mmHg. However, the pressure sensitivity still needs to be improved to satisfy clinical diagnostic requirements. Here, we investigated the effects of acoustic parameters on the pressure sensitivity of microbubbles through measuring the acoustic attenuation and scattering properties of commercially available SonoVue microbubbles. Our results showed that the first harmonic, subharmonic, and ultraharmonic amplitudes of microbubbles were reduced by 6.6 dB, 10.9 dB, and 9.3 dB at 0.225 mechanical index (MI), 4.6 dB, 19.8 dB, and 12.3 dB at 0.25 MI, and 18.5 dB, 17.6 dB, and 12.6 dB at 0.3 MI, respectively, when the ambient pressure increased from 0 to 180 mmHg. Our finding revealed that a moderate MI (0.25⁻0.4) exciting microbubbles could significantly improve their sensitivities to detect ambient pressure.
Related Concept Videos
Detection of Gross Error: The Q Test
Detection of Black Holes
Their closest cousins are neutron stars, which are composed almost entirely of neutrons packed against each other, making them extremely dense. A neutron star has the same mass as the Sun but its diameter is only a few kilometers. Therefore, the escape velocity from their surface is close to the speed of light.
Not until the 1960s, when the first neutron...
Difference from Background: Limit of Detection
The LOD indicates the presence or absence...
Effects of EDTA on End-Point Detection Methods
In the visual method, metal-ion indicators (metallochromic dyes), which have distinct colors in their free and complex forms, are added to the mixture to signal the titration's end point. They form stable complexes with metal ions, but these complexes are weaker than the corresponding metal–EDTA complexes. As a...
Types of Errors: Detection and Minimization
Absolute error in a measurement is the numerical difference from the true or central value. Relative error is the ratio between absolute error and the true or central value, expressed as a percentage.
Errors can be classified by source, magnitude, and sign. There are three types of errors: systematic, random, and gross.
Systematic or...
Precipitation Titration: Endpoint Detection Methods
In the Volhard method, a standard excess of AgNO3 is first added to the...