Technical Note: The design, construction, and evaluation of a liquid-based single phantom solution for TG128
Brian Leong1, Mark Ostyn1, Seungjong Oh1
1Department of Radiation Oncology, Virginia Commonwealth University, Richmond, VA, 23298, USA.
Purpose:
Since the publication of the AAPM TG128 report for the quality assurance (QA) of prostate brachytherapy ultrasound systems, no commercially available phantoms have been developed which satisfy all of the task group recommendations. Current solid phantoms require a separate user-implemented setup using a container with liquid medium to evaluate the alignment between the needle template and the electronic grid, a test of geometric accuracy with critical implications in dosimetric quality. Utilizing a 3D printer, we constructed a cost-effective, liquid-based phantom that provides a complete TG128 solution which improves the efficiency of brachytherapy ultrasound QA.
Methods:
The TG128 report was used to guide the design process of the liquid-based phantom. The needle template and electronic grid alignment setup served as the foundation with specific components developed to integrate all remaining tests. Water was chosen as the liquid medium, with speed of sound adjusted to 1,540 m/s via salinity per the task group recommendations. The proof of concept was evaluated by comparing the time stamps labeled on QA images between the liquid-based phantom and a commercially available one for both a new and experienced user.
Results:
A TG128 QA trial run demonstrated that all recommended tests can be completed with the single phantom setup. Evaluation of the time data revealed a total QA duration of 45 min (average of two trials) with the liquid-based phantom, compared to 70 and 90 min with the commercial phantom for a new and experienced user.
Conclusions:
The liquid-based phantom is specifically designed to satisfy the recommendations of the TG128 report. The incorporation of 3D printing allows simple design modifications to adapt the phantom on-the-fly if needed. The resulting product improves the efficiency of brachytherapy ultrasound QA by eliminating the need for multiple phantom setups.
More Related Videos
06:33Construction of a Preclinical Multimodality Phantom Using Tissue-mimicking Materials for Quality Assurance in Tumor Size Measurement
Published on: July 29, 2013
09:32Manufacturing Abdominal Aorta Hydrogel Tissue-Mimicking Phantoms for Ultrasound Elastography Validation
Published on: September 19, 2018
Related Concept Videos
Weak Base Solutions
Strong Acid and Base Solutions
Design Example: Dimensioning of Concrete Masonry Construction
The site engineer has laid out a plan for the storeroom with external dimensions of twelve feet in length and...
Solution Formation
This selective...
Leveling Effect and Non-Aqueous Acid-Base Solutions
The Leveling Effect of a Solvent
A generic acid (HA) reacts with the generic base (B-) to yield the corresponding conjugate base (A-) and conjugate acid (HB):
Solution Composition During Acid/Base Titrations
The α0 and α1 values...