Impact of statistical reconstruction and compressed sensing algorithms on projection data elimination during X-ray CT
Bing-Yu Sun1, Yoshihiko Hayakawa2
1Course of Medical Engineering, Graduate School of Engineering, Kitami Institute of Technology, 165 Koencho, Kitami, Hokkaido, 090-8507, Japan.
Objectives:
To examine the effect of incomplete, or total elimination of, projection data on computed tomography (CT) images subjected to statistical reconstruction and/or compressed sensing algorithms.
Methods:
Multidetector row CT images were used. The algebraic reconstruction technique (ART) and the maximum likelihood-expectation maximization (ML-EM) method were compared with filtered back-projection (FBP). Effects on reconstructed images were studied when the projection data of 360° (360 projections) were decreased to 180 or 90 projections by reducing the collection angle or thinning the image data. The total variation (TV) regularization method using compressed sensing was applied to images processed by the ART. Image noise was subjectively evaluated using the root-mean-square error and signal-to-noise ratio.
Results:
When projection data were reduced by one-half or three-quarters, ART and ML-EM produced better image quality than FBP. Both ART and ML-EM resulted in high quality at a spread of 90 projections over 180° rotation. Computational loading was high for statistical reconstruction, but not for ML-EM, compared with the ART. TV regularization made it possible to use only 36 projections while still achieving acceptable image quality.
Conclusions:
Incomplete projection data-accomplished by reducing the angle to collect image data or thinning the projection data without reducing the angle of rotation over which it is collected-made it possible to reduce the radiation dose while retaining image quality with statistical reconstruction algorithms and/or compressed sensing. Despite heavier computational calculation loading, these methods should be considered for reducing radiation doses.
Related Concept Videos
Reconstruction of Signal using Interpolation
Radiological Investigation I: X-ray and CT
Bioequivalence Data: Statistical Interpretation
Statistical Methods for Analyzing Epidemiological Data
X-ray Imaging
Electron Microscope Tomography and Single-particle Reconstruction
Electron Tomography
Electron tomography can be performed either in TEM or STEM (scanning transmission...