Magnitude dependent discordance in liver stiffness measurements using elastography point quantification with
Ivica Grgurevic1, Nermin Salkic2, Tonci Bozin3
1Department of Gastroenterology, Hepatology and Clinical Nutrition; Department of Medicine, University Hospital Dubrava, University of Zagreb School of Medicine and Faculty of Pharmacy and Biochemistry, Avenija Gojka Suska 6, 10000, Zagreb, Croatia. ivica.grgurevic@zg.htnet.hr.
Point shear wave elastography (ElastPQ) shows promise for assessing liver fibrosis in chronic liver diseases. However, ElastPQ and transient elastography (TE) measurements are not interchangeable, particularly at higher stiffness values.
Area of Science:
- Hepatology
- Medical Imaging
- Diagnostic Performance Evaluation
Background:
- Chronic liver diseases (CLD) necessitate accurate non-invasive methods for staging liver fibrosis.
- Transient elastography (TE) is a common method, but point shear wave elastography (ElastPQ) offers an alternative approach.
Purpose of the Study:
- To evaluate the diagnostic performance of ElastPQ for non-invasive liver fibrosis assessment in CLD patients.
- To compare ElastPQ with TE as a reference standard for liver stiffness measurement (LSM).
Main Methods:
- Liver stiffness was measured using both TE and ElastPQ in 165 CLD patients and 20 healthy volunteers.
- Fibrosis staging by TE served as the reference standard.
- Statistical analyses included correlation, ROC analysis, Bland-Altman plots, and Passing-Bablok regression.
Main Results:
- ElastPQ demonstrated high diagnostic performance for fibrosis stages F≥2 (AUC=0.96), F≥3 (AUC=0.98), and F4 (AUC=0.98).
- While TE and ElastPQ measurements correlated significantly (r=0.86), they were not interchangeable, especially when TE values exceeded 10 kPa.
- Bland-Altman and Passing-Bablok analyses revealed significant differences between the two methods at higher stiffness levels.
Conclusions:
- ElastPQ is a reliable method for liver fibrosis assessment, particularly relevant for non-alcoholic fatty liver disease (NAFLD).
- LSM values from TE and ElastPQ are not interchangeable, with discrepancies increasing above 10 kPa.
- Further research is warranted to explore ElastPQ's utility in sub-classifying compensated advanced chronic liver disease.
Related Concept Videos
Orders of Magnitude
The order of magnitude is simply a way of rounding numbers consistently to the nearest power of 10. This makes doing rough mental math...
Inertial Frames of Reference
Non-inertial Frames of Reference
Chemical Shift: Internal References and Solvent Effects
The internal reference compound generally used in NMR spectroscopy is tetramethylsilane (TMS). TMS is preferred because it is chemically inert, soluble in NMR solvents, and easily removable. Also, the highly shielded methyl protons in TMS yield an intense...
Series R—L Circuit Transients
Using Kirchhoff's Voltage Law (KVL) to analyze this circuit helps determine the total asymmetrical fault current, which consists...
Frequency-dependent Selection