Radiation dose of coronary CT angiography with a third-generation dual-source CT in a "real-world" patient population
Aleksander Kosmala1, Bernhard Petritsch1, Andreas Max Weng1
1Department of Diagnostic and Interventional Radiology, University Hospital Würzburg, Oberdürrbacherstrasse 6, 97080, Würzburg, Germany.
Objectives:
To assess radiation dose and image quality of coronary computed tomography angiography (cCTA) with a third-generation dual-source scanner in a real-world patient population.
Methods:
Scans of otherwise unselected, consecutive patients referred for clinically indicated cCTA between June 2015 and November 2017 were included for retrospective analysis. Scan protocol was based on heart rate: prospectively ECG-gated high-pitch spiral below 60 beats per minute (bpm), prospectively ECG-gated sequential scan between 61 and 70 bpm, and retrospective spiral above 70 bmp or at irregular heart rates. Objective image quality was measured as signal-to-noise (SNR) and contrast-to-noise ratio (CNR); subjective image quality was evaluated using a five-point Likert scale by two independent readers. For radiation dose analysis, effective dose, size-specific dose estimates, and volume CT dose index were assessed.
Results:
Two hundred seventy-eight patients (median age, 60 years; 155 men) with a median body mass index of 26.6 kg/m2 (range, 16.7-60.9 kg/m2; 180 (64.7%) overweight or obese) were included (122 in the high-pitch spiral group, 60 in the prospective sequence group, and 96 in the retrospective spiral group). Median effective dose was 0.63 mSv (interquartile range [IQR], 0.51-0.90 mSv) for high-pitch spiral, 1.32 mSv (IQR, 0.79-2.46 mSv) for prospective sequence, and 4.77 (IQR, 3.02-8.27 mSv) for retrospective spiral (p < 0.001). Most studies had at least very good image quality (91.4/88.8% R1/R2), with highest SNR and CNR in the high-pitch spiral group.
Conclusions:
cCTA with sufficient image quality is achievable at reasonably low radiation exposure in a real-world patient collective with a high proportion of overweight or obese patients.
Key Points:
• Submillisievert radiation dose coronary CT angiography with good diagnostic image quality is feasible in the majority of cases in a real-world patient using high-pitch spiral. • Prospective sequence results in about double median effective dose compared to the high-pitch protocol. • To optimize individual radiation exposure, lowering the heart rate is paramount, as it allows for choosing a dose-optimized (high-pitch spiral) scan protocol.
More Related Videos
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Radiological Investigation I: X-ray and CT
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
Imaging Studies for Cardiovascular System V: CT
Imaging Studies for Cardiovascular System VI: Calcium -Scoring CT
Imaging Studies III: Computed Tomography