Real-time simultaneous shim and motion measurement and correction in glycoCEST MRI using double volumetric navigators
Gizeaddis L Simegn1, Andre J W Van der Kouwe1,2,3, Frances C Robertson1,4
1UCT Medical Imaging Research Unit, Division of Biomedical Engineering, Department of Human Biology, University of Cape Town, Cape Town, South Africa.
Purpose:
CEST MRI allows for indirect detection of molecules with exchangeable protons, measured as a reduction in water signal because of continuous transfer of saturated protons. CEST requires saturation pulses on the order of a second, as well as repeated acquisitions at different offset frequencies. The resulting extended scan time makes CEST susceptible to subject motion, which introduces field inhomogeneity, shifting offset frequencies and causing distortions in CEST spectra that resemble true CEST effects. This is a particular problem for molecules that resonate close to water, such as hydroxyl group in glycogen. To address this, a technique for real-time measurement and correction of motion and field inhomogeneity is proposed.
Methods:
A CEST sequence was modified to include double volumetric navigators (DvNavs) for real-time simultaneous motion and shim correction. Phantom tests were conducted to investigate the effects of motion and shim changes on CEST quantification and to validate the accuracy of DvNav motion and shim estimates. To evaluate DvNav shim and motion correction in vivo, acquisitions including 5 experimental conditions were performed in the calf muscle of 2 volunteers.
Results:
Phantom data show that DvNav-CEST accurately estimates frequency and linear gradient changes because of motion and corrects resulting image distortions. In addition, DvNav-CEST improves CEST quantification in vivo in the presence of motion.
Conclusion:
The proposed technique allows for real-time simultaneous motion and shim correction with no additional scanning time, enabling accurate CEST quantification even in the presence of motion and field inhomogeneity.
Related Concept Videos
Real Time RT-PCR
The real-time quantification of the number of amplified products is...
Distance Corrections
Power Factor Correction
Equation of Motion: General Plane motion
Moreover, the body's center of mass experiences a rotational effect as a result of these couple moments. This rotation can be articulated as the...
Fixing Double-strand Breaks
NMR Spectrometers: Resolution and Error Correction