Durability Data for Bioprosthetic Surgical Aortic Valve: A Systematic Review
Benish Fatima1, Divyanshu Mohananey1, Fazal W Khan2
1Department of Cardiovascular Medicine, Heart and Vascular Institute, Cleveland Clinic, Cleveland, Ohio.
Importance:
Although several studies have reported data on surgical aortic valve durability, variability in study methodologies and definitions as well as inadequate follow-up make the interpretation of data from these studies difficult to interpret.
Objective:
To review available data on structural valve deterioration (SVD) of surgical bioprosthetic aortic valves by examining the published literature as well as data reported to the US Food and Drug Administration (FDA).
Evidence Review:
A systematic review using Preferred Reporting Items for Systematic Review and Meta-Analysis reporting guidelines to obtain all available data from preexisting literature on the actuarial freedom from SVD and outcomes of SVD. Data were collected from database inception to November 2016. Additionally, data reported to the FDA were obtained.
Findings:
In total, 167 studies and 12 FDA reports including 101 650 patients and 17 different valve types were analyzed. There were 11 different definitions of SVD used in the studies. Core laboratory data were available for 11 studies. Mean follow-up ranged from less than 1 year to 14 years. Kaplan-Meier estimates were reported for up to 30 years of follow-up, with 0% to 37% of patients remaining at risk at maximum follow-up. Food and Drug Administration-reported data are similarly variable in duration (mean follow-up range, 1.2 to 7 years) and completeness of follow-up, with final follow-up of 0.1% to 95%.
Conclusions And Relevance:
There is considerable variability in reporting SVD of surgical aortic valves, with different definitions and inadequate long-term systematically collected core laboratory data. Rigorously collected long-term data with standardized definitions for surgical valves are needed to provide a benchmark for the durability of rapidly evolving transcatheter valves.
Related Concept Videos
Review and Preview
Percentiles are a type of fractile that partition data into...
Review and Preview
Brick Durability, Strength, and Appearance
Heart Valves
The AV valves prevent the backflow of blood from the ventricles to the atria during ventricular contraction. These valves function with the assistance of the chordae tendineae and papillary muscles. When the ventricles are relaxed, the chordae tendineae are slack, allowing blood to flow from the atria into the...
Random and Systematic Errors
Systematic Sampling Method
Systematic sampling is one of the simplest methods...