Sensitivity analysis of neurodynamic and electromagnetic simulation parameters for robust prediction of peripheral
Valerie Klein1,2, Mathias Davids1,3, Lawrence L Wald3,4,5
1Computer Assisted Clinical Medicine, Medical Faculty Mannheim, Heidelberg University, Heidelberg, Germany.
Abstract:
Peripheral nerve stimulation (PNS) has become an important limitation for fast MR imaging using the latest gradient hardware. We have recently developed a simulation framework to predict PNS thresholds and stimulation locations in the body for arbitrary coil geometries to inform the gradient coil optimization process. Our approach couples electromagnetic field simulations in realistic body models to a neurodynamic model of peripheral nerve fibers. In this work, we systematically analyze the impact of key parameters on the predicted PNS thresholds to assess the robustness of the simulation results. We analyze the sensitivity of the simulated thresholds to variations of the most important simulation parameters, including parameters of the electromagnetic field simulations (dielectric tissue properties, body model size, position, spatial resolution, and coil model discretization) and parameters of the neurodynamic simulation (length of the simulated nerves, position of the nerve model relative to the extracellular potential, temporal resolution of the nerve membrane dynamics). We found that for the investigated setup, the subject-dependent parameters (e.g. tissue properties or body size) can affect PNS prediction by up to ~26% when varied in a natural range. This is in accordance with the standard deviation of ~30% reported in human subject studies. Parameters related to numerical aspects can cause significant simulation errors (>30%), if not chosen cautiously. However, these perturbations can be controlled to yield errors below 5% for all investigated parameters without an excessive increase in computation time. Our sensitivity analysis shows that patient-specific parameter fluctuations yield PNS threshold variations similar to the variations observed in experimental PNS studies. This may become useful to estimate population-average PNS thresholds and understand their standard deviation. Our analysis indicates that the simulated PNS thresholds are numerically robust, which is important for ranking different MRI gradient coil designs or assessing different PNS mitigation strategies.
Related Concept Videos
Sensitivity, Specificity, and Predicted Value
Sensitivity is the...
The Electromagnetic Spectrum
The Electromagnetic Spectrum
Peripheral Nervous System: Ganglia and Nerves
Nerves
The nerve is a bundle of axons that serves as the communication highway in the PNS. Each nerve is ensheathed in a protective layer of connective tissue called the epineurium. This outermost layer safeguards the nerve and supports the...
Local Anesthetics: Differential Sensitivity of Nerve Fibers
Noncompartmental Analysis: Miscellaneous Pharmacokinetic Parameters
One key aspect of the noncompartmental approach is determining a drug's total clearance. This can be done by dividing the drug dose by the area under the concentration-time curve from zero to infinity. The area under the concentration-time curve represents the drug's...