Development and dosimetric assessment of a patient-specific elastic skin applicator for high-dose-rate brachytherapy
So-Yeon Park1, Seonghee Kang2, Jong Min Park3
1Department of Radiation Oncology, Veterans Health Service Medical Center, Seoul, Republic of Korea; Institute of Radiation Medicine, Seoul National University Medical Research Center, Seoul, Republic of Korea.
Purpose:
The purpose of this study was to develop a patient-specific elastic skin applicator and to evaluate its dosimetric characteristics for high-dose-rate (HDR) brachytherapy.
Methods And Materials:
We simulated the treatment of a nonmelanoma skin cancer on the nose. An elastic skin applicator was manufactured by pouring the Dragon Skin (Smooth-On Inc., Easton, PA) with a shore hardness of 10A into an applicator mold. The rigid skin applicator was printed using high-impact polystyrene with a shore hardness of 73D. HDR plans were generated using a Freiburg Flap (FF) applicator and patient-specific rigid and elastic applicators. For dosimetric assessment, dose-volumetric parameters for target volume and normal organs were evaluated. Global gamma evaluations were performed, comparing film measurements and treatment planning system calculations with various gamma criteria. The 10% low-dose threshold was applied.
Results:
The V120% values of the target volume were 56.9%, 70.3%, and 70.2% for HDR plans using FF, rigid, and elastic applicators, respectively. The maximum doses of the right eyeball were 21.7 Gy, 20.5 Gy, and 20.5 Gy for the HDR plans using FF, rigid, and elastic applicators, respectively. The average gamma passing rates were 82.5% ± 1.5%, 91.6% ± 0.8%, and 94.8% ± 0.2% for FF, rigid, and elastic applicators, respectively, with 3%/3 mm criterion.
Conclusions:
Patient-specific elastic skin applicator showed better adhesion to irregular or curved body surfaces, resulting in better agreement between planned and delivered dose distributions. The applicator suggested in this study can be effectively implemented clinically.
More Related Videos
06:20Irradiator Commissioning and Dosimetry for Assessment of LQ α and β Parameters, Radiation Dosing Schema, and in vivo Dose Deposition
Published on: March 11, 2021
06:39Electroencephalographic, Heart Rate, and Galvanic Skin Response Assessment for an Advertising Perception Study: Application to Antismoking Public Service Announcements
Published on: August 28, 2017
Related Concept Videos
Drug Dosing: Geriatric Patients
Drug Dosing: Obese Patients
Application of Rates of Change
Dose-Response Relationship: Selectivity and Specificity
Drug Dosing in Renal Diseases: Dose Adjustments Based on Drug Clearance and Elimination Rate Constant
Assessment of Ventilation I: Respiratory Rate
A Ventilation assessment is critical for monitoring a patient's health status. Respiration, one of the most accessible vital signs, provides insights into the function of numerous body systems and can indicate serious health issues, such as brainstem injuries from head trauma.
Critical Guidelines for Assessing Ventilation: