Spinal phantom comparability study of Cobb angle measurement of scoliosis using digital radiographic imaging
Ni Chung1, Yi-Hong Cheng1, Hiu-Lam Po1
1Department of Health Technology and Informatics, The Hong Kong Polytechnic University, Hong Kong, China.
Background:
Computed radiography (CR), digital radiography (DR) and biplanar radiography (EOS™ imaging system) are common imaging tools for radiographic evaluation of adolescent idiopathic scoliosis (AIS). The effect of imaging methods in relation to later-on Cobb angle measurements on radiographs is not yet quantified. The study aimed to examine the compatibility between CR, DR and EOS for scoliotic quantification by evaluating the reliability, agreement of different imaging methods, and assessing the prediction performance for EOS measurement from that of CR and DR.
Method:
A flexible spine phantom was used to simulate 32 different scoliotic curves ranging from 10° to 60°. Each curvature was imaged using DR, CR and EOS systems accordingly. Each of the six observers independently measured Cobb angle twice on each image at a two-week interval. Intraclass correlation coefficient (model 2 and 3), Bland-Altman plot and linear regression analysis were completed to evaluate the reliability, agreement, and the prediction of Cobb angle measurement, respectively.
Results:
Reliability analysis showed excellent intra-observer reliability (Intraclass correlation coefficient >0.9) for each observer and good inter-observer reliability (Intraclass correlation coefficient = 0.84 for EOS; 0.739 for CR; 0.877 for DR) for each method. Bland-Altman plots demonstrated good agreement between imaging methods without fixed or proportional bias. Excellent coefficient of determination was achieved, with 0.980 for CR versus EOS measurements, and 0.973 for DR versus EOS measurements.
Conclusions:
Radiographs produced by all of the three methods can provide reliable and accurate Cobb angle measurements for scoliosis assessments. None of the methods systemically underestimates or overestimates the Cobb angle measurement. Additionally, all of the evaluated methods are satisfactory in obtaining images for Cobb angle measurement in AIS. However, the 3D post-processing techniques offered by EOS should also be taken into consideration as it takes a vital role in treatment and monitoring of 3D deformity in the case of scoliosis.
The Translational Potential Of This Article:
In view of the limited availability of biplanar radiography (EOS™ imaging system), computed radiography and digital radiography are demonstrated to be reliable alternatives in scoliosis monitoring as evident in the reliability, agreement and prediction of Cobb angle measurement.
More Related Videos
Related Concept Videos
Contact Angle
The adhesive force is the molecular force between molecules of different materials, that is, between the molecules of the solid and the liquid. The cohesive...
Open Angle Glaucoma: Treatment
Drugs such as carbonic anhydrase inhibitors, α2- and...
Angle of Twist: Problem Solving
Comparing Copy Number Variations and SNPs
Copy number variations or CNVs are the structural variations that cover more than 1kb of DNA sequence. The single nucleotide polymorphism (SNP), on the other hand, is a single nucleotide change or a point mutation that is found in more than 1%...
Imaging Studies VII: Vascular Imaging
Imaging Studies II: Ultrasonography