MICRODOSIMETRY AT NANOMETRIC SCALE WITH AN AVALANCHE-CONFINEMENT TEPC: RESPONSE AGAINST A HELIUM ION BEAM
D Mazzucconi1,2, D Bortot1,2, S Agosteo1,2
1Politecnico di Milano, Dipartimento di Energia, via La Masa 34, Milano, Italy.
Abstract:
The tissue-equivalent proportional counter (TEPC) is the most accurate device for measuring the microdosimetric properties of a particle beam but, since the lower operation limit of common TEPCs is ~0.3 μm, no detailed information on the track structure of the impinging particles can be obtained. The pattern of particle interactions at the nanometric level is measured directly by only three different nanodosimeters worldwide: practical instruments are not yet available. In order to partially fill the gap between microdosimetry and track-nanodosimetry, a low-pressure avalanche-confinement TEPC was designed and constructed for simulating tissue-equivalent sites down to the nanometric region. The present paper aims at describing the response of this TEPC in the range 0.3 μm-25 nm to a 62 MeV/n 4He ion beam. The experimental results, for depths near the Bragg peak, show good agreement with FLUKA simulations and suggest that, for smaller depths, the distribution is highly influenced by secondary electrons.
Related Concept Videos
pH Scale
Ion Channels
Ion channels are specialized integral membrane proteins on the plasma membrane that allow...
Beams
Based on geometry, beams can be straight, tapered, or curved. Straight beams are the most common type and have a constant cross-section throughout their length. Tapered beams, on the other hand, have a varying cross-section along...
Common Ion Effect
Precipitation of Ions
The equation that describes the equilibrium between solid calcium carbonate and its solvated ions is:
Formation of Complex Ions