Are metal artefact reduction algorithms effective to correct cone beam CT artefacts arising from the exomass?
Amanda Pelegrin Candemil1, Benjamin Salmon2, Deborah Queiroz Freitas1
11 Department of Oral Diagnosis, Division of Oral Radiology, Piracicaba Dental School, University of Campinas , Piracicaba , Brazil.
Abstract:
The aim of this study was to evaluate the effectiveness of metal artefact reduction (MAR) in cone beam CT (CBCT) artefacts arising from metallic objects in the exomass. A radiographic phantom composed of 16 polypropylene tubes filled with a homogeneous radiopaque solution was created. CBCT scans were obtained with two units: Picasso Trio (Vatech, South Korea) and ProMax (Planmeca, Finland). The phantom was centred in a 5 × 5 cm field-of-view (FOV) with titanium and CoCr inserts in the exomass. All scans were repeated after enabling MAR. Mean voxel values were obtained from the 16 tubes and standard deviation was calculated as a way of measuring voxel value variability. Mean values and voxel value variability were compared individually in the presence and absence of MAR by means of analysis of variance, followed by Tukey's test (α = 0.05). In the Picasso Trio, MAR significantly decreased mean voxel values (p ≤ 0.05) and increased voxel value variability (p > 0.05) in the presence of titanium. When CoCr was present, no statistical difference (p > 0.05) was observed. In the ProMax, MAR increased significantly mean voxel values (p ≤ 0.05) in the presence of titanium, and presented no significant difference (p > 0.05) for CoCr. Voxel value variability did not differ significantly (p > 0.05) for both materials. In conclusion, MAR was not effective to correct CBCT artefacts arising from metallic objects in the exomass in the two CBCT units used.
More Related Videos
09:49Dual-phase Cone-beam Computed Tomography to See, Reach, and Treat Hepatocellular Carcinoma during Drug-eluting Beads Transarterial Chemo-embolization
Published on: December 2, 2013
10:25Single-Digit Nanometer Electron-Beam Lithography with an Aberration-Corrected Scanning Transmission Electron Microscope
Published on: September 14, 2018
Related Concept Videos
Oxidation-Reduction Reactions
Metal-Ligand Bonds
In these complexes, transition metals form coordinate covalent bonds, a kind of Lewis acid-base interaction in which both of the electrons in the bond are contributed by a donor (Lewis base) to an electron acceptor (Lewis acid). The Lewis acid in...
Distance Corrections
Beams
Based on geometry, beams can be straight, tapered, or curved. Straight beams are the most common type and have a constant cross-section throughout their length. Tapered beams, on the other hand, have a varying cross-section along...
Power Factor Correction
Bonding in Metals